Download Vortex - Invacare
Transcript
1. B Charging problems If the battery or charger develop any faults during charging, the green charging light and the reconditioning light on the charger and/or the battery’s capacity lights will start to flash. 1. The charging light and the reconditioning light flash simultaneously (figure 1): • Faulty battery. Replace the battery with a new one. A 2. C B 2. The fourth capacity light on the battery starts to flash (figure 2): • Faulty battery. Replace the battery with a new one. The charging light (A) and reconditioning light (B) flash at one second intervals: • A • This may be a contact fault. Check whether there are dust or other particles between the charger and the lead. A fuse may have blown. Open the fuse box on the side of the battery and check that everything works. If not, replace the blown fuse with the spare one stored on the side of the battery cover. Warning! Use the indicated fuses only (30A fuse flat) to avoid the risk of the battery catching fire. 24 VORTEX