Transcript
Data Log
The Data Log sub menu contains three options:
•
•
•
Run Manual
Run Prog 1
Run Prog 2
Use the ▲ ▼ keys to make a selection, and press ↵ to accept.
Run Manual immediately starts a data logging session (called a
"Test").
Press ↵ or PAGE to stop a test in progress. The AM510 will ask you to
confirm that you want to stop the test or not, while data logging
continues. If you choose to continue (Return) by pressing the PAGE
key, there will be no break in the data.
While running a manual test, the display shows LOGGING DATA as
well as the last data value recorded. Manual tests use the logging
interval defined in Log Interval under the Setup Menu.
Each test you run is assigned a sequential number (e.g., TEST#1,
TEST#2, … TEST#100) until memory is used up or cleared.
When a test is stopped, the display immediately switches to the
Statistics Menu where you can view the following statistics for any
stored test:
•
•
•
•
•
Maximum reading
Minimum reading
Average reading
8-hour TWA
Time (duration of test)
First, select the test to view with the ▲ ▼ keys and press ↵. Now, you
can page through the available statistics using the keys. Initially, the test
you just completed is offered. You can view test statistics any time you
are not actively data logging by selecting Statistics under the Main
Menu.
Run Prog 1 and Run Prog 2 allow you to execute a data logging test
that was preprogrammed using TRAKPRO Software. Preprogramming
allows a great deal of flexibility that is not available with manual
logging such as predefined start and stop times, logging intervals,
calibration factors, and more. See “Programming Prog 1 and Prog 2
with TRAKPRO Software” later in this chapter.
20