appropriate ranges manually by yourself. The difference between
auto-range and manual-range is the settling time. Auto-range is a
convenient way for user, but manual-range can usually speed up the
process.
If the input signal is beyond allowed range, an “OVLD” message will be
shown on the display. The thresholds of maximum/minimum readings
for each range are 120 % of the range for maximum and 10% of the
range for minimum.
Default
The default is auto-range. The user selected range is stored in volatile
memory and the default will be restored when the meter is power-off.
Please refer to Table 2-1 on page 24 for the factory default range.
How to set the auto / manual range:
You can set the auto / manual range either through the front panel
operation or through the remote interface operation.
Front Panel Operation
First choose a measurement function on the front panel, then press
“AUTO” button to select auto-range feature. Or use △ and ▽ buttons
to manually select a range.
If selected range is not shown on the
bottom of the display, press round PREV or NEXT buttons until the range
information appears.
The locations of these buttons are shown with a
red rectangle frame in Figure 4-6.