Transcript
Techniques and Applications Manual experiment potential limits. This will considerably reduce the noise level and increase the plot quality. 3.2.2 CstV: Constant Voltage The constant voltage (CstV) technique is especially dedicated to supercapacity testing. It is designed to apply successively several voltage steps to the cell(s). Between each voltage step, an open circuit voltage period can be added. Rest period The rest period is an open circuit voltage period. Refer to the OCV description for more details. Potential step with data recording conditions Apply Ei = … V vs. Ref/Eoc/Ectrl/Emeas. the potential step is defined vs. Ref the reference electrode potential or according to the previous open circuit potential (Eoc), controlled potential (Ectrl) or measured potential (Emeas). for ti = … h … mn … s sets the potential step duration. Limits |I| to IMax = … pA/…/A and |Q| < QM = ... fA.h/…/A.h/pC/…/kC. Imin = … pA/…/A curtails the step duration if the current or charge limit is reached. If the limit is reached, the loop condition (go to Ns' for nc times), if set, is not used, and the program continues to the next sequence (Ns + 1). The |Q| value is the integral charge for the current sequence. This value is not reset if there is a loop on the same sequence (Ns' = Ns). 0 values disable the tests. Record I every dIp = … pA/…/A, dQp = ... fA.h/…/A.h/pC/…/kC and dtp = … s <I> every dt = … s you can record either an instantaneous current value I or an averaged current value <I>. The recording conditions during the potential step depend on the chosen current variable. For the instantaneous current the recording values can be entered simultaneously. Then it is the first condition reached that determines the recording. A zero value disables the recording for each criterion. For the averaged current the user defines the time for the average calculation. In that case the data points are recorded in the channel board memory every 200 µs for VMP3 based instruments and for VMP300 based instruments. Leave dI alone for Chronoamperometry experiments, and dQ for Chronocoulometry experiments. 167