USER MANUAL FLOWANALYSER
6.4.3 RS 232
TM
The RS232 interface is used for servicing (firmware download),
TM
for the connection of the MultgasAnalyser OR-703 as well as
for remote control of the unit and is located on the back of the
TM
FlowAnalyser .
RS 232
The connection to the RS 232 port has to be established over
the special RS 232 Download Cable (>12.4 Spare Parts)
If the instrument has to be remote controlled e.g. through a
specific software package you can as your authorized dealer for
a detailed description of the protocol.
Assignment:
The pin numbers are equal on the instrument as well as on the
D-Sub 9 plug where as pin 1 on the instrument is on the top.
Pin 5
Pin 7
Pin 8
Pin 1,2,3,4,6,9
Version 2.1
GND
TxD
RxD
Not connected