IC CONTROLS Sensor Instructions Chemical Cleaning Sensor Storage Chlorine sensors can be refreshed with p/n A1100227, Chlorine sensor renew solution. Short Term: Immerse the sensor tip in tap water. This procedure should be done over a sink. Wear plastic or rubber gloves as the solution is acidic. Wash hands thoroughly with lots of water if the solution comes in contact with the skin. 1. Remove the protective cap exposing the coils and gold tip. 2. Immerse in cleaning solution as shown (Figure 19) for 5 to 10 minutes, or until deposits Wet storage is good for a week or two. CAUTION: If a wet sensor dries out in storage there may be none repairable damage. LongTerm: Dis-assemble the Chlorine sensor tip and pour out the fill solution. Rinse the coils, gold tip, and membrane module with demin water and blot dry with a paper towel. Re-assemble the Chlorine sensor dry, and store dry with the tip covered. Dry storage can be used for a year or more. D.O. sensor beaker Keep coils below surface Figure 19 Chlorine Sensor Cleaning disappear. 3. Remove and rinse in Distilled or Demin water, use wiping in the direction of the coils or a toothbrush to speed removal. 4. Repeat 2 & 3 until coils and tip look clean and have a shine, then re-assemble Chlorine sensor with new membrane and fill solution, calibrate and check efficiency is above 50 %. 5. Repeat 2 to 4 as necessary to get at least 50 percent efficiency. If not possible, Chlorine sensor should be replaced. Page 26 www.iccontrols.com 875 Instructions