Download Clarus 600/650 D MS Hardware Guide
Transcript
Clarus 600/560 D MS Hardware Guide Compressed Gases Compressed Gases. High pressure gas cylinders can be dangerous if mishandled or misused. Always handle gas cylinders with caution and observe your local regulations for the safe handling of gas cylinders. WARNING Avoid banging the valves, and ensure that the correct valves and gauges are installed. It is recommended that gas cylinders be stored and placed outside the laboratory and connected to the instrument through specially cleaned copper tubing. Take care not to kink or stress the gas tubing. For safety, cylinders must be firmly clamped in an upright position. Explosive hazard. When using hydrogen, methane or isobutane, special care must be taken to avoid buildup of explosive gas mixtures either in the GC oven or the mass spectrometer vacuum manifold. WARNING Ensure that all hydrogen line couplings are leak-free and do not allow hydrogen to vent within the oven. Ventilation WARNING 22 Hazardous vapors. When analyzing hazardous compounds, such as pesticides, or running in the chemical ionization (CI) mode, it is necessary to vent the mass spectrometer effluent from the forepump exhaust into a fume hood or charcoal trap.