Download Manual
Transcript
BATTERY RECHARGING The wheelchair batteries are recharged by connecting the charger plug into the driving controller’s charger socket. During recharging, the battery charge level is shown as an icon on the display and the wheelchair will not be possible to drive. Batteries must be recharged immediately, if the icon indicating the battery charge level shows only two full pillars. Follow the instructions supplied with the charger. Use only the charger that has been supplied with the wheelchair. Using incorrect chargers could damage the batteries, wheelchair or the charger itself. FUSES The wheelchair’s main fuse (80A) is situated between the batteries. If it is blown, the wheelchair will not have any electrical functions. The user must not change this fuse themselves; the wheelchair must be taken for service. R-NET www.chasswheel.com 21