Download Product Specifications
Transcript
Wahl C150 Calibrator C150 Calibration Functions and performances: @23°C +5°C Accuracy is given in % of reading (C150 display) + fixed value DC Current: Emission DC Current: Measurement Range Measuring Range Resolution Accuracy (1 year) Remarks Range Resolution Accuracy (1 year) 0mA/24mA 0.1µA 0.007% + 0.8µA Rin < 30Ω 24mA 1µA 0.007% R + 0.8µA 4-20mA 3mA/24mA 0.1µA 0.007% + 0.8µA Rin < 30Ω 4-20mA 1µA 0.007% R + 0.8µA 100mA 0mA/100mA 0.1µA 0.009% + 2µA Rin < 30Ω 0-20mA 1µA 0.007% R + 0.8µA 0-20mA C150 can measure up to 100mA with/without loop supply (24V). Loop supply = 24 V ±100% Temperature coefficient: < 10% of accuracy /°C. Display unit: °C, and °F HART compatibility: input impedance Rin=280Ω Display with linear or quadratic scaling Pre-programmed steps 0% 25% 50% 75% 100% 4-20mA linear 4 8 12 16 20 0-20mA linear 0 5 10 15 20 4-20mA quad 4 5 8 13 20 0-20mA quad 0 1.25 5 11, 25 20 4-20mA valves Measuring Range Resolution Accuracy (1 year) +100mV 10mV + 100µV 3.8 - 4 - 4.2 12 19, 20, 21 DC Voltage: Emission DC Voltage: Measurement Range Emission output with or without 24V loop supply 1µV 0.005% + 2µV Range Remarks Meas Range Resolution Accuracy (1 year) Min Load 1µV 0.005% R + 2µV 1KΩ -15mV + 1V 10µV 0.005% R + 8µV 2KΩ +100mV -5mV + 100mV Rin > 10MΩ +1V -100mV + 1V 10µV 0.005% + 8µV Rin > 10MΩ +1V +10V -1V + 10V 100µV 0.007% + 80µV Rin = 1MΩ +10V -100mV + 10V 100µV 0.007% R + 80µV 4KΩ +50V -5V + 50V 1mV 0.007% + 0.5mV Rin = 1MΩ +50V -100mV + 50V 1mV 0.007% R + 0.5mV 4KΩ Rin = Input resistance Frequency and Counting: Measurement Frequency and Pulses: Emission Range Resolution Accuracy (1 year) Range Resolution Accuracy (1 year) 10kHz < 0.01 Hz 0.01% Rdg 1000 Hz < 0.01 Hz 0.01% R 100kHz 0.1 Hz 0.01% 100 kHz 1 Hz 0.01% R Threshold triggering: 1V Unit scale: Pulse/min or Hz Measurement on frequency signals and on dry contacts Measurement for counting will be performed on defined time or infinite time. Unit scale: Pulse/min or Hz Pulse emission Dry contacts simulation Max amplitude: 20V (Selectable by user) Resistance: Measurement Resistance: Emission Range Input Range Resolution Accuracy (1 year) Remarks Accuracy (1 year) Remarks 400Ω 0 to 400Ω 1mΩ 0.006% R + 8mΩ 4 wires 400Ω 1 -4000Ω 10mΩ 0.006%R +8mΩ @1mA 3600Ω 0 to 3600Ω 10mΩ 0.006% R + 50mΩ 4 wires 4000Ω 1-3600Ω 100mΩ 0.006%R +50mΩ @0.1mA I ext of 0.1 to 1mA 50KΩ 0 to 50KΩ 100mΩ 0.008% R + 1Ω 4 wires 50KΩ 1-50KΩ 1Ω 0.008%R +1Ω @1mA Range Output Range Resolution Resistance measurement with 2, 3 or 4 wires resistance measurement: automatic recognition of number of connected wires displayed on the screen. Pressure: measurement with external digital sensor Range 0-1 bar 0-3 bar 0-10 bar 0-30 bar 0-100 bar 0-300 bar 0-1000 bar Absolute Yes Yes Yes Yes Yes Yes Yes Relative Yes Yes Yes Yes N/A N/A N/A Resolution: 0.02% of Full Scale Accuracy: 0.05% of Full Scale for 10°C and 40°C, -0.1% of Full Scale -10°C to +10°C and 40°C to 80°C. I ext of 5µA to 50µA Emission with pulsed available: refer to the manual for specifications Temperature coefficient: < 5ppm /°C. 0°C to 18°C and 28°C to 50°C Emission resistance: establishing time < 1ms: for compatibility with smart transmitters WARRANTY Calibration Services Available (800) 421-2853 • FAX (828) 658-0728 • www.palmerwahl.com 07/10 Rev A 3 Continued Innovation Since 1836 ISO 9001:2008 CERTIFIED