Download Project Deliverable
Transcript
Summary Test Priority Level of expertise* Output repeatability Output reproducibility Filtration Essential Essential Essential MPE MPE MPE Tube potential Field size and alignment Leakage Essential Essential Essential MPE MPE MPE Quantitative image Quality Image density values Uniformity and artifacts Recommended Essential Display specific Noise Limiting resolution Contrast detail Geometrical accuracy General condition Patient dose X-ray tube and generator Suggested frequency Action levels** Mean ±10% Baseline ±10% < 2.5mm aluminium (of which 1.5mm should be permanent) In house/MPE In house 12 monthly 12 monthly When new, if output changes or tube head dismantled 12 monthly 12 monthly When new and if damage suspected Monthly Monthly Recommended Essential Recommended Essential Essential In house/MPE In house/MPE In house/MPE In house/MPE In house 12 monthly 12 monthly 12 monthly 12 monthly Monthly Monitor resolution Patient dose index Recommended Recommended In house MPE Monthly 12 monthly Patient dose audit Essential In house/MPE At least 3 yearly > ±5% of intended kV >10% expected field size > 1000µGy hr-1 at maximum tube rating. >10% from baseline Visible artefacts on the image or >±10% of the mean > ±10% from baseline > ±20% from baseline Dependent on method used. within ±2mm and ±2º Failure to resolve different contrasts in test pattern/ not consistent between monitors Not consistent with baseline image Outside ±15% of manufacturer’s specification > national or international reference level Notes * Level of expertise: MPE indicates that this test would normally require the input of a medical physics expert with sophisticated test equipment whereas in house indicates that the tests can normally be performed by clinic staff using standard phantoms ** Action level: Results outside these levels should be investigated and action taken. The advice of a medical physics expert or service engineer may be required N.B. This table represents initial guidance based on current experience of dental CBCT units. It should be kept under critical review as experience is gained in testing such units. 33 SEDENTEXCT D3.4 Report – Appendix 1: Generic QA Protocol