SOP No.: 406
Page 8 of 9
Environmental Quality Lab
Chemistry Procedures
Rev. 3
3-15-09
Table 1. Quality Control Requirements for Turbidity by Hach 2100N Turbidimeter
QC Sample or Activity
Minimum Frequency
Acceptance Criteria
Capability demonstration
Four (4) prepared
samples analyzed prior to
any customer sample
analyses
Criteria for duplicate precision
4-Point calibration plus
deionized water blank
At least quarterly
90-110% R (measured value of primary
standards when analyzed as samples
within 10% of expected values)
Assign values to
permanent transfer
standards using formazin
primary standards
At least quarterly
Measurement after acceptable 4-point
calibration and values within 10% of
previous established values
Daily calibration check
Immediately prior to and
after sample analysis
GELEX Secondary Turbidity Standards
should read within 10% of assigned values
Method blank
Daily prior to sample
analysis
<1.0 NTU (i.e., < RL)
Sample analysis
For all sample analyses
Direct sample reading within acceptable
measurement range (i.e., 1.00 NTU to
4000 NTU)
Sample duplicate
One (1) per preparation
batch
RPD < 25%
Internal PE sample
Blind PE sample
LCS =
MB =
MDL =
MS =
PE =
Samples and frequency
determined by Lab QA
Officer
Samples and frequency
determined by
accrediting agencies and
projects
laboratory control sample
method blank
method detection limit
matrix spike
performance evaluation
QC =
%R =
RL =
RPD =
Corrective Action
Repeat until acceptable
Investigate problem. Correct
any obvious problems. Repeat
calibration until acceptable.
Investigate problem. Correct
any obvious problems
including replacing transfer
standards if necessary. Repeat
until acceptable.
Investigate problem. Correct
any obvious problems. If
necessary reassignment of
GELEX values and reanalyze
samples. Repeat calibration
check until acceptable.
Clean analytical system and
repeat MB analysis. Identify
and eliminate source of
contamination.
If reading below range report
result as < RL. If result above
range dilute sample.
Investigate problem. If system
precision is in control, qualify
results. If system precision is
out of control, reanalyze entire
batch.
75-125% R
RPD < 25%
Investigate all unacceptable
results.
Determined by PE provider
Investigate all unacceptable
results.
quality control
percent recovery
reporting limit
relative percent difference