Download ARU-700
Transcript
SYMPTOMS POSSIBLE CAUSE PROBABLE SOLUTION Low or no cell current. Cathode not securely fastened to power head. Securely tighten cathode coupling screw to power head. ERR 1/LO-A flashing on display. Set screw holding motor coupler to cathode coupling has loosened. Tighten set screw in motor coupler. Cathode contact brush worn out. Unit has been ON for too long a period of time. All of the silver has been removed from fixer. Recovery unit ON too long. Replace cathode contact brush. Lower set point and or current settings. Recovery from bleach-fix. Unit won't recover silver from bleach fix. Cathode contaminated with grease and or oil. Sulphiding (rotten egg smell coming from silver recovery unit) Silver falling off cathode. Check electrical source No silver collecting on cathode. Solution is not fixer. Clean cathode with clean cloth and scouring powder (comet, etc.) and water; rinse thoroughly in water to remove contaminates. Use silver laden fixer. ON/OFF switch ON, no power. No silver in fixer. Blown fuse. Use silver laden fixer. Investigate and remedy cause, replace fuse. 21