Download User`s Manual
Transcript
Section 9 Optical System Check Press ◄ or ► from the Setup Screen to select the Optical Check mode, as shown below. Note: • The control device is intended to check the optical system. • Allow the control devices and the meter to reach operating temperature (15-40°C or 59-104°F) prior to testing. • The optical check should be performed under normal lab lighting conditions. Do not perform under sunlight or extreme lighting conditions. to enter this mode. Press as shown below. The meter will flash the test device symbol, Insert a control device into the Device Channel. Follow the direction of the arrows indicated on the device. Ensure that the control device is inserted all the way. Press normal. to start the optical check. If the meter displays YES, the meter is If the meter displays no, the meter is not functioning properly. 36