Download Data Sheet - Vox Power

Transcript
MEDICAL D A T A S H E E T NEVO
O+600
0 SERIIES
‐ Modular ‐ Configurab
ble ‐ Reduced faan noise ng ‐ Enhanced LLine Deratin
600 wa
att in the pa
alm of your hand
enhanced NEVVO+ series featuures improved low input line derating deliveers 600 watts w
without the neeed to derate foor line voltages above The Vac making thee NEVO+ the ultimate power solution for products destined for the glob
bal market. Through the implementation off a fan 120V
trol circuit the NEVO+ also features reducedd fan noise. Thhe NEVO+ delivvers up to 600 watts from a 6600 gram 5” x 3” x 1U packagge and cont
conssists of an inpu
ut module toge
ether with up tto eight isolate
ed output modules. The serie
es carry full ULL60601 3rd ED ssafety approva
als and comply with EN610000‐3, EN610000‐4 and EN5500022‐B EMC stanndards. N FEATURES: MAI
 



 No Line deratin
N
ng above 120VA
AC R
Reduced fan no
oise 6
600 watt outpu
ut 5
5” x 3” x 1U foo
otprint 3
H
High power den
nsity (25W/in ) )
rd
U
UL 60601 3
ED
D 










High reliabillity High efficien
ncy – up to 89%
% Only 0.6kg –– 100W/kg
I2C control o
option
Remote current/voltage programming Current output signal Accurate current sharing 5V
V 200mA bias su
upply Ro
oHS compliant Fie
eld configurable
e Tw
wo year warranty SYSTEM
M SPECIFICATIO
ONS
INSTALLATION I
INP
PUT ELECTRICALL Param
meter AC Input Voltage AC Input Frequency DC Input Voltage Powe
er Rating Details e is 100Vrms to Nominal range
240Vrms Contact factorry for 400Hz operation. Medical Inrush
h Current See graphs for deratings 600Watts output at 120Vrms input 265Vrms (cold
d start) Fusing 5x20 Fast acting Inputt Current Limit Maintains pow
wer factor Inputt Current Min 85 47 Typ
50/
60
120 Max 264 63 Units
Hz II IP R
Rating IP10
0 Pollution d
degree 2 ROH
HS Com
mpliance
2002
2/95/EC Material ggroup IIIb (indoorr use only) 600 Wattss 6 Ampss 20 Ampss 8 Ampss RELIABILITY Ampss Componen
nt Details Mag Lev Std
d Excluding FA
AN See individu
ual output datasheets See graphs 86 % Idle P
Power All outputs fittted and enabled 28 Wattss INPUT Idle P
Power 21 Wattss OUTPUT Warranty
UVLO
O All outputs fittted and Disabled Typical value ffor 300Watts output at 240Vrms input 600Watts output at 120Vrms input Turn on only Over temperature Internally mon
nitored. Latching Reliab
bility 40°C 80% load
d Holdu
up Bias Voltage S i g n a l s Bias current 17 mS 78 84 Vrms
115 125 °C 20 2 4.8 5 0 PNP open collector with internal 10k pull down
n resistor 8 Inhibit voltage Inhibit current 10k ohm input impedance Min Max Units 2.7 FPMH 2 FPMH 1 FPMH 2 Years 21 Power Good voltage Power Good current Global inhibit voltage Global inhibit current 0.99 94V‐‐2 I Installation
n category
Efficie
ency 0.96
Detaails Equipment class FAN Powe
er Factor Parrameter Vrms
Vdc 89 Details Flam
mmability ratiing 300 8 Parameterr 5.2 200 10 SAFETY Parameterr FPMH
H Isolation Voltage V mA Isolation Clearance
Details Max Units Input to Ou
utput Min 4000 Vac Input to Ch
hassis 1500 Vac Output to Chassis 250 Vdc Output to Output Primary to Secondary (Reinforced) Primary to Chassis (Basic) Primary to Secondary (Reinforced) Primary to Chassis (Basic) 250 Vdc 15 V 0 20 mA 2 15 V Isolation Creepage
0.2 1.5 mA Leakage Cu
urrent Medical: 265Vac, 63Hz, 25°C 7 mm 2.5 mm 12 4 mm 300 5k ohm input impedance 3 15 V 0.6 3 mA Parameterr Details MECHANICAL AC_OK voltage 1 4 V Size 33.7mm x 41.0mm 77.7mm x 13
(all external dimensions ± 1.0mm) AC_OK current ‐10 20 mA Weight
360 gram +6
60 gram per output module
AC_OK warningg See user manu
ual for exceptions mS Mounting
Bottom or Siide mounting (See d
diagram for details)) Vox Power Ltd. Unitt 2 Red Cow Interrchange Estate Ballyymount Dublin 24, Ireland 3 5511 Tel: +353 1 459 11611, Fax: +353 1 633
b: ww.vox‐powerr.com Web
5 Distributed By: mm uA MEDICAL DATASHEET
D A T A S H E E T NEVO
N O+600 SERIES
EMC
E
Sttandard Leve
el Radiated electtric field EN
N55011, EN55022, FFCC B Conducted em
missions EN
N55011, EN55022, FFCC B Harmonic Distortion EN
N61000‐3‐2 Com
mpliant m
m Flicker & Fluctuation 106 kP
Pa Electrostatic discharge EN
N61000‐3‐3 N61000‐4‐2 EN
(1
15kV air, 8kV contacct) Min Max U
Units Temperature ‐40 +85 °C
C Humidity Relative, non‐con
ndensing 5 95 %
% Altitude ‐200 5000 Air Pressure 54 Full power ‐20 50 °C
C Temperature Derate input and outputs at 2.5%/°C 50 70 °C
C Humidity Relative, non‐con
ndensing 5 95 %
% Altitude (‐200 to 2000m fo
or UL60601‐1) Air Pressure ‐200 3000 m
m 78 106 kP
Pa 35 60 dBA Noise Level Variable. Measured 1m from fan intake Shock 3000 bumps at 10
0G (16ms) half sine wave Vibration 1.5G 10 to 200Hz sine wave, 20G forr 15min in 3 axes random vibration
n Emissions
Emissions Parameter
Details Immunity
Immunity O p e r a t i o n S t o r a g e ENVIRON
NMENTAL Parameter Com
mpliant 4 Radiated RFI EN
N61000‐4‐3 (10V/m
m) 3 Fast Transient burst EN
N61000‐4‐4 (4kV) 4 Input line surgges EN
N61000‐4‐5 (1kV L‐N
N, 2kV L‐E) 3 Conducted RFII EN
N61000‐4‐6 (10V) 4 Power Freq. M
Magnetic Field EN
N61000‐4‐8 (10A/m
m) 3 Voltage Dips EN
N61000‐4‐11 (EN55
5024) Com
mpliant AGENCY APPROVALS
A
Sta
andard File IEC
C 60601‐1: 1998 + A1: 1991 + A2: 1995 86 UL: E31648
UL6
60601‐1: 2006 CAN
N/CSA‐C22.2 No. 600601‐1 (2008) ANSI/AAMI ES60601‐11 (2005 + C1:09 + A2
2:10) CE MARK ‐ LVD 2011/665/EU CB certificate and repo
ort available on req
quest 600
550
500
450
400
350
300
250
80
100
120
140
160
180
Vinrms
200
220
240
260
Power out
Power out
Temperature Derating
g
650
600
Line Dera
ating
650
550
500
450
400
350
300
250
-30
-10
-
30
Tem
mperature
Vox Power Ltd. Unitt 2 Red Cow Interrchange Estate Ballyymount Dublin 24, Ireland 3 5511 Tel: +353 1 459 11611, Fax: +353 1 633
b: ww.vox‐powerr.com Web
10
Distributed By: 50
70
9
90
MEDICAL DATASHEET
D A T A S H E E T NEVO
N O+600 SERIES
MEECHANICAL DIM
MENSIONS AND
D MOUNTING SSCREWS N - Neatral
E - Earth
L - Live
SC
CREWS
MH1, MH
H2, MH3, MH4, MH5
5 Screw typ
pe M4 Tightening Penetration depth Tighten
n to 1.5 Nm 4.00mm max including chassis OUTPUT M
MODULES x 8 Screw typ
pe M3x5, C/Sink, Posi, Stainleess Steel Tightening Penetration depth Tighten
n to 0.75 Nm Defined by screw CHASSIS xx 5 Screw typ
pe M3x5, C/Sink, Posi, Stainleess Steel
Tightening Penetration depth Tighten
n to 0.75 Nm Defined by screw FAN x 2 Screw typ
pe M3x30
0, C/Sink, Posi, Stain
nless Steel Tightening Penetration depth Tighten
n to 0.75 Nm SLO
OT D
Defined by screw SLO
OT C
Airflow
SLO
OT B
SLO
OT A
Vox Power Ltd. Unitt 2 Red Cow Interrchange Estate Ballyymount Dublin 24, Ireland 3 5511 Tel: +353 1 459 11611, Fax: +353 1 633
b: ww.vox‐powerr.com Web
Distributed By: MEDICAL D A T A S H E E T DATASHEET
NEV
VO600
0+ SER
RIES
CONNECTOR
RS
J J1
PINOUTS J1 Circuit 1 2 3 Details Live Earth Neutrall J2 Circuit 1 2 3 4 5 6 7 8 9 10 11 12 Circuit 1 2 3 4 5 6 REF. Details Power G
Good Slot A Inhibit Power G
Good Slot B Inhibit Power G
Good Slot C Inhibit Power G
Good Slot D Inhibit Global Inhibit AC OK +5V 2000mA Bias Supply COM J5 Details ‐Sense +Sense Voltagee Control Currentt Control / Share / O
Out COM +5V loccal bias supply J3 Positive Output J2
J4 Circuit # 2
Circuit # 1
Negative Output +
-
DET
TAILS TERMINAL
M
MANUFACTUR
RER
J1 M
MOLEX
10013036
MAIN
NS INPUT: 3 Pin, 5.0
08mm, with Friction Lock, 18‐24 AWG
J2 M
MOLEX
511101260
0 GLOB
BAL SIGNALS: 12 Pin
n, 2mm, with Frictio
on Lock, 24‐30 AWG
G
J3/4(1) V
VARIOUS
OUTP
PUT POWER TERMIN
NAL: TAB SIZE 6.35m
mmx0.8mm J5 M
MOLEX
OUTP
PUT SIGNALS: 6 Pin,, 1.25mm, with Fricttion lock, 28‐32 AW
WG 051021060
00 Notes 1. Terminal and W
Wire current rating must exceed maxim
mum short circuit ou
utput current. Eg. O
Output 1 = 25A*1.25
5 = 31.25Amps 2. Direct equivaleents may be used fo
or any connector paarts 3. All cables mustt be rated 105°C min, equivalent to UL1015 0008701031
0503948051
VARIOUS 0500588000 PART
T NUMBERING
G SYSTEM
NEVO+ Power series Leakagee Current NEVO+ 600
0 M ‐ 1 1 2 3
3 ‐ 0 0 0 Factory Use Use ‘0’ for un
nused slots. Blanking platees will be M – Meedical inserted at factory Slot A –– Output # Slot D – Output #
Slot B –– Output # Slot C – Output # When initially o
ordering non‐nomin
nal voltage settings add “ /Voltage” aftter each output # w
where a special voltaage setting is requirred E.g. If 3.30Vdc is required in slo
ot B and all other slo
ots require nominal voltages then use:: NEVO+ 600M – 1 – 1/3..30 – 2 ‐ 3
hat can be used for aall future orders of the same configuraation
The factory will tthen issue a 3 digit code for your speciific configuration th
When ordering an input unit with no outputs in
nserted, simply order NEVO+ 600M Vox Power Ltd
d. reserves the right to change or improve any part of the specificattion, electrical or mech
hanical design or manu
ufacturing process with
hout notice. Please con
nsult your local distribu
utor or contact Vox Pow
wer to ensure that you have the latest specification
n before using your pro
oduct. For other inform
mation relating to the u
use of the product pleaase refer to the latest N
NEVO user manual. Vo
ox Power reserves the rright to make changes without notice to ducts. Vox Power does not assume any liabilitty arising out of the use or application of anyy of its products and off any information to the maximum extent perrmitted by law. No liceense, express or implied, by estoppel or any of its prod
otherwise, to aany intellectual properrty rights is granted by this document or by aany conduct of Vox Pow
wer. VOX POWER DISCLAIMS ALL WARRANTIES AND REPRESENTATIONS. IN PARTICULAR A
ALL OTHER WARRANTIES, CONDITION OR TERMS REL
LATING TO SUITABILITYY, FITNESS FOR PURPO
OSE, MECHANTABILITY OR CONDITION OF THE PRODUCTS AND WHETHER EXPRESS OR IMPLIED BY STATURE OR COMMON LAW OR OTTHERWISE ARE EXCLUD
DED. Document DO
OC6008 rev 04 Vox Power Ltd. Unit 2 Red Cow Inteerchange Estate Ballymount Dublin 24, Ireland 33 5511 Tell: +353 1 459 11661, Fax: +353 1 63
Weeb: ww.vox‐poweer.com Distributed
d By: