Dual Port
Differential
Manometer
Accessory Head
Model: ADMN2
Description
P1
P2
Dual-Port
Manometer
Accessory
Head
0.1
ON
Lo Batt
ENGLISH
Resolution
The ADMN2 is a portable dual port manometer
accessory head. The ADMN2 is capable of taking
gas pressure as well as static pressure. The
ADMN2 also has a zero adjust. The auto-power-off
function conserves battery life, but can be disabled
for data logging if desired.
The ADMN2 will allow you to take gas pressure
between -60inWC and 60inWC. The ADMN2 can
measure P1 or P2 pressure readings or the difference between P1 and P2. The ADMN2 also has two
different measurement scales: inWC (inches of
water column) and mBar. Static pressure is possible
with the resolution of 0.01inWC or 0.01mBar.
0.01
METRIC
(LCD/10)
Operation
P1-P2
P2
P1
ZERO
AUTO
OFF
ADMN2
OPERATOR’S MANUAL
Specifications
Pressure Ports: Two tube connectors for 4mm
(3/16 inch) I.D. flexible tubing
Units of Measure: inWC and mbar
Accuracy:
±0.02inWC, ±(0.00inWC to 2.00inWC);
±1.5% FS, ±(2.00inWC to 60.0inWC);
±0.05mbar, ±(0.00mbar to 5.00mbar);
±1.5% FS, ±(5.00mbar to 150.0mbar);
Resolution: 0.1 or 0.01 depending on resolution
selected (If 0.01 resolution is selected you must
divide displayed value by 10 for actual value).
Measurement Range:
inWC: -60.0 to 60.0
mBar: -150.0 to 150.0
Operating environment:
32ºF (0°C) to 122ºF (50°C), under 75%RH
Compatible media: Dry, non-corrosive gases
Battery: Single standard 9-volt battery, NEDA
1604, JIS 006P, IEC 6F22.
Battery Life: 100 hours with low battery LED
indicator on accessory head.
Overrange: “OL” or “-OL” is displayed.
Auto power off: 15 minutes
Dimensions: 131mm(~5 3/16”)(H) x 67mm(~2
7/16”)(W) x 32mm(~1 1/4”)(D)
Weight: approximately 154g (~1/2 pound)
including battery
1. Connect to COM and Volts jacks using any
Fieldpiece test leads with removable probe tips.
For Fieldpiece “stick-style” meters, EHDL1, or
DL3, slide the head directly over the meter.
2. Select the mVDC range.
3. Zero the ADMN2 by pressing the ZERO button
just before taking the pressure readings, while
at ambient pressure. This will zero both P1 and
P2.
4. Connect a single hose to get the gauge pressure relative to the ambient or ZERO pressure.
5. Connect both hoses if you want to see relative
pressure, P1 minus P2.
Resolution Modes (0.1 and 0.01)
There are two modes of resolution on the
ADMN2. For most applications, 0.1 mode will be
fine to use. When measuring low pressures, 0.01
will likely be the best mode for meaningful testing.
The most common circumstance for 0.01 mode is
measuring static pressure across a blower.
6. Select P1, P2, or P1 - P2.
7. Select English (inches of water column) or
Metric (mBar) units.
8. To disable the auto-power-off, remove the rubber covering and flip the switch below. Autopower-off is helpful for conserving the battery;
but if you are data logging with Fieldpiece model
DL3, you will want to disable this feature.
9. If you are in an environment where the temperature is noticeably changing while you are taking
your reading, disconnect the meter from the
hoses and ZERO it relative to ambient pressure
before each reading.
Checking Gas Pressure on a Regulator
1. Screw the brass fitting into the pressure port on
the regulator.
2. Put unit into operation (i.e. turn on the furnace
and have furnace ignite, as if running it in normal operation.)
3. This will give you the pressure coming out of the
regulator.
4. If you suspect high or low inlet pressure into the
regulator, the manometer can hook into the inlet
port in the same manner it can connect into the
outlet port. If you have a dual-port manometer,
you can check both the inlet and the outlet
simultaneously and see the pressure drop
across the regulator.
5. See manufacturer's specification for the specifiOnly when using 0.01 mode, the display of your
meter will show the true measurement X 10. So if
the true measurement is 0.08, your meter will show
0.8. This is done to optimize resolution.
0.1
0.1
P1
0.01
(LCD/10)
P1
P2
Dual-Port
Manometer
Accessory
Head
0.1
ON
0.01
(LCD/10)
P2
Dual-P
Dual-Port
Ma
Manom
e
Manometer
Acces
cessory
Accessory
Head
0.1
ON
Lo Batt
0.05
inWC
ENGLISH
E
Resolution
tion
Resolution
ZERO
AUTO
OFF
ZERO
AUTO
Remove the rubber covering the ZERO button.
By pressing the ZERO button, both the P1 and P2
are zeroed to the ambient pressure. For this reason
the calibration should be done when both P1 and
P2 are disconnected from the hoses in order to
properly calibrate to ambient pressure.
Connect Your Way
P1
P1
P1
P1
P2
P2
0.1
ON
Lo Batt
ENGLISH
Resolution
Dual-Port
Manometer
Accessory
Head
0.1
ON
Lo Batt
ENGLISH
Resolution
0.01
METRIC
(LCD/10)
Dual-Port
Manometer
Accessory
Head
0.1
ON
Lo Batt
ENGLISH
0.01
METRIC
(LCD/10)
Dual-Port
Manometer
Accessory
Head
0.1
ON
Lo Batt
ENGLISH
Resolution
0.01
METRIC
(LCD/10)
P1-P2
P2
P1
P1-P2
P2
P1
Resolution
0.01
METRIC
(LCD/10)
ZERO
ZERO
P1-P2
P2
P1
P1-P2
P2
P1
P2
P2
Dual-Port
Manometer
Accessory
Head
AUTO
AUTO
OFF
OFF
ADMN2
ADMN2
ZERO
ZERO
AUTO
AUTO
OFF
OFF
ADMN2
ADMN2
OFF
2000
EHDL1
LT series
DL3
HS series
Warranty
The product is warranted to the original purchaser against defects in material or workmanship for a
period of one (1) year from the date of purchase.
During the warranty period, Fieldpiece Instruments
will, at its option, replace or repair the defective unit.
This warranty does not apply to defects resulting
from abuse, neglect, accident, unauthorized repair,
alteration, or unreasonable use of the instrument.
Any implied warranty arising out of the sale of
Fieldpiece's products including but not limited to
implied warranties of merchantability and fitness for
purpose are limited to the above. Fieldpiece shall
not be liable for incidental or consequential damages.
Return any defective ADMN2 to Fieldpiece for
warranty service along with proof of purchase.
Contact Fieldpiece for out of warranty repair
charges.
P1-P2
P2
P1
0.01
METRIC
(LCD/10)
P1-P2
P2
P1
Field Pressure Calibration
Service
0.01
01
METRIC
ME
(LCD/10)
Lo Batt
ENGLISH
cations on target inlet and outlet pressure for a
given regulator or piece of combustion equipment.
ADMN2
OFF
3.0 inWC
ADMN2
To Meter
To Meter
3.0 mVDC
Gas pressure measurement
(0.1 resolution mode)
0.8 mVDC
-0.03
inWC
Static pressure across a blower
(0.01 resolution mode)
0.8mVDC / 10 = 0.08inWC
v13
RMA316 Manometer Adapter
Used for 5/16" pressure outlet ports.
1. Shut off main gas supply to furnace.
2. Move the gas valve switch to "OFF" position.
3. Use a 3/32" hex wrench to loosen the outlet
pressure tap screw. Rotate screw counterclockwise one revolution to open.
4. Connect the ADMN2 tubing to the smaller
(3/16") end of the adapter and the other (5/16")
end of the adapter into the of the adapter tube.
5. Slide the 5/16" adapter tube over the outlet
pressure boss (port) to seal. Overlap the pressure boss by at least 3/8" to prevent leakage.
Smaller end (3/16”) of the adapter
connects to ADMN2 tubing. Remove
the installed brass fitting from tube
so the adapter can be inserted.
Adapter
(5/16” to 3/16”)
Adapter Tube
(5/16”)
Outlet Pressure
Tap Screw
Inlet Pressure
Tap Screw
Outlet
Pressure
Boss
Inlet Pressure
Boss
More Products From Fieldpiece
CLAMP
300V
CAT.III
400A
AC Current
Clamp
Stick Meter
This is the heart of modular expandability. In addition to being a full functioning multimeter, any accessory head can be used with it.
!
1AAC / 1mVAC
400AAC MAX
ACH4
A
& ir
He Te Velo
ad mp cit
era y
tu
re
ON
LE
AB
ST
n
a tiot g
lin
er
rig rhea bcoo 34A
R1
R ef pe Su
Su d
an
LO
TT
BA
EN
GL
H
IS
SH
En
glish
Ft
/m
M
M in
et
PH
ric
ºF
KM M/s
OFF
/h
r
ºC
IC
TR
ME
SC
A
04
R4
SH
SC p
or m re
Te
su
Pr
es
t
Se M
AT
T/C l
Ca
RE
AD
ag
se e
c)
4
AA
V3
AU
OFFTO
-
T/C
CAL AU
TO
OFF
2
AOX2
°F
TEMP
%CO
%O 2
°C
ON
Lo Batt
GAS
LC
10 D X
0
BA
TT
Model HS36
Re
tim al
e
OIL
BLE
Fue
l Typ
e
STA
ON
LO
is
h Av
(16 er
ric
Accessory heads are the sensors of multiple
parameters measured by technicians every day.
They plug into a mV range (depending on the head)
of a multimeter. The multimeter will display whatever the head is measuring. Instead of having to purchase and carry a separate instrument for each
parameter, a technician can use multiple heads and
a single multimeter to do the job.
TO
AUOFF
X2
AS
Comb
Chec ustio
n
k
En
gl
M
et
Accessory Heads
Modular Expandability
Modular expandability is ability for accessory
heads and meters to change configurations to
match the various needs of an HVAC/R technician.
Accessory heads (the sensors) send out a mV
signal, which represents the value of the measurement, to whatever meter is attached to it. Heads can
attach directly to the top of a Stick meter, DL3 data
logger, or EHDL1. They can also plug into any meter
with mV ranges using ASLS2 leads.
Non contact voltage
Magnetic hanger
Autoranging
Backlight
Temperature
Volts, amps, ohms
Frequency
Microfarads
aaaaaaa
aaaaa
aaaaa
a a a
aaaa
a
aa
aa
a
aa
aaa
aaaaa
LCD X
100
READ
ON
Combustion
Check
STABLE
Lo Batt
GAS
Fuel Type
Metric
English
Metric
Ft/min
MPH
ºF
M/s
KM/hr
ºC
OIL
Real
time
TEMP
%CO2
%O2
OFF
AUTO OFF
AUTOOFF
AAV3
Includes:
HS36 Meter
ACH4 Current Clamp
ATB1 K-type TCouple
ADLS2 Deluxe Leads
ANC1 Case
ON
LO BATT
Air Velocity
& Temperature Average
English (16 sec)
Head
ARH4
°C °F
AOX2
Here are four of the many heads available:
AAV3 Air Velocity and Temperature
ARH4 Digital Psychrometer
AVG2 Digital Vacuum Gauge
AOX2 Combustion Check
T/C AUTO
CAL
OFF
ASP2 Static Pressure Probe
For accurate static pressure measurements.
1. Connect ASP2 to ADMN2 via hose.
2. Insert ASP2 into drilled or pre-existing 1/4” hole.
3. Use allignment arrow on the ASP2 to direct
probe into the air stream.
4. Read static pressure measurements from “stickmeter”, EHDL1 or record with DL3.
5. Remove probe and patch hole.