Download ADMN2 - Dual Port Manometer Accessory Head
Transcript
Dual Port Differential Manometer Accessory Head Model: ADMN2 Description P1 P2 Dual-Port Manometer Accessory Head 0.1 ON Lo Batt ENGLISH Resolution The ADMN2 is a portable dual port manometer accessory head. The ADMN2 is capable of taking gas pressure as well as static pressure. The ADMN2 also has a zero adjust. The auto-power-off function conserves battery life, but can be disabled for data logging if desired. The ADMN2 will allow you to take gas pressure between -60inWC and 60inWC. The ADMN2 can measure P1 or P2 pressure readings or the difference between P1 and P2. The ADMN2 also has two different measurement scales: inWC (inches of water column) and mBar. Static pressure is possible with the resolution of 0.01inWC or 0.01mBar. 0.01 METRIC (LCD/10) Operation P1-P2 P2 P1 ZERO AUTO OFF ADMN2 OPERATOR’S MANUAL Specifications Pressure Ports: Two tube connectors for 4mm (3/16 inch) I.D. flexible tubing Units of Measure: inWC and mbar Accuracy: ±0.02inWC, ±(0.00inWC to 2.00inWC); ±1.5% FS, ±(2.00inWC to 60.0inWC); ±0.05mbar, ±(0.00mbar to 5.00mbar); ±1.5% FS, ±(5.00mbar to 150.0mbar); Resolution: 0.1 or 0.01 depending on resolution selected (If 0.01 resolution is selected you must divide displayed value by 10 for actual value). Measurement Range: inWC: -60.0 to 60.0 mBar: -150.0 to 150.0 Operating environment: 32ºF (0°C) to 122ºF (50°C), under 75%RH Compatible media: Dry, non-corrosive gases Battery: Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22. Battery Life: 100 hours with low battery LED indicator on accessory head. Overrange: “OL” or “-OL” is displayed. Auto power off: 15 minutes Dimensions: 131mm(~5 3/16”)(H) x 67mm(~2 7/16”)(W) x 32mm(~1 1/4”)(D) Weight: approximately 154g (~1/2 pound) including battery 1. Connect to COM and Volts jacks using any Fieldpiece test leads with removable probe tips. For Fieldpiece “stick-style” meters, EHDL1, or DL3, slide the head directly over the meter. 2. Select the mVDC range. 3. Zero the ADMN2 by pressing the ZERO button just before taking the pressure readings, while at ambient pressure. This will zero both P1 and P2. 4. Connect a single hose to get the gauge pressure relative to the ambient or ZERO pressure. 5. Connect both hoses if you want to see relative pressure, P1 minus P2. Resolution Modes (0.1 and 0.01) There are two modes of resolution on the ADMN2. For most applications, 0.1 mode will be fine to use. When measuring low pressures, 0.01 will likely be the best mode for meaningful testing. The most common circumstance for 0.01 mode is measuring static pressure across a blower. 6. Select P1, P2, or P1 - P2. 7. Select English (inches of water column) or Metric (mBar) units. 8. To disable the auto-power-off, remove the rubber covering and flip the switch below. Autopower-off is helpful for conserving the battery; but if you are data logging with Fieldpiece model DL3, you will want to disable this feature. 9. If you are in an environment where the temperature is noticeably changing while you are taking your reading, disconnect the meter from the hoses and ZERO it relative to ambient pressure before each reading. Checking Gas Pressure on a Regulator 1. Screw the brass fitting into the pressure port on the regulator. 2. Put unit into operation (i.e. turn on the furnace and have furnace ignite, as if running it in normal operation.) 3. This will give you the pressure coming out of the regulator. 4. If you suspect high or low inlet pressure into the regulator, the manometer can hook into the inlet port in the same manner it can connect into the outlet port. If you have a dual-port manometer, you can check both the inlet and the outlet simultaneously and see the pressure drop across the regulator. 5. See manufacturer's specification for the specifiOnly when using 0.01 mode, the display of your meter will show the true measurement X 10. So if the true measurement is 0.08, your meter will show 0.8. This is done to optimize resolution. 0.1 0.1 P1 0.01 (LCD/10) P1 P2 Dual-Port Manometer Accessory Head 0.1 ON 0.01 (LCD/10) P2 Dual-P Dual-Port Ma Manom e Manometer Acces cessory Accessory Head 0.1 ON Lo Batt 0.05 inWC ENGLISH E Resolution tion Resolution ZERO AUTO OFF ZERO AUTO Remove the rubber covering the ZERO button. By pressing the ZERO button, both the P1 and P2 are zeroed to the ambient pressure. For this reason the calibration should be done when both P1 and P2 are disconnected from the hoses in order to properly calibrate to ambient pressure. Connect Your Way P1 P1 P1 P1 P2 P2 0.1 ON Lo Batt ENGLISH Resolution Dual-Port Manometer Accessory Head 0.1 ON Lo Batt ENGLISH Resolution 0.01 METRIC (LCD/10) Dual-Port Manometer Accessory Head 0.1 ON Lo Batt ENGLISH 0.01 METRIC (LCD/10) Dual-Port Manometer Accessory Head 0.1 ON Lo Batt ENGLISH Resolution 0.01 METRIC (LCD/10) P1-P2 P2 P1 P1-P2 P2 P1 Resolution 0.01 METRIC (LCD/10) ZERO ZERO P1-P2 P2 P1 P1-P2 P2 P1 P2 P2 Dual-Port Manometer Accessory Head AUTO AUTO OFF OFF ADMN2 ADMN2 ZERO ZERO AUTO AUTO OFF OFF ADMN2 ADMN2 OFF 2000 EHDL1 LT series DL3 HS series Warranty The product is warranted to the original purchaser against defects in material or workmanship for a period of one (1) year from the date of purchase. During the warranty period, Fieldpiece Instruments will, at its option, replace or repair the defective unit. This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument. Any implied warranty arising out of the sale of Fieldpiece's products including but not limited to implied warranties of merchantability and fitness for purpose are limited to the above. Fieldpiece shall not be liable for incidental or consequential damages. Return any defective ADMN2 to Fieldpiece for warranty service along with proof of purchase. Contact Fieldpiece for out of warranty repair charges. P1-P2 P2 P1 0.01 METRIC (LCD/10) P1-P2 P2 P1 Field Pressure Calibration Service 0.01 01 METRIC ME (LCD/10) Lo Batt ENGLISH cations on target inlet and outlet pressure for a given regulator or piece of combustion equipment. ADMN2 OFF 3.0 inWC ADMN2 To Meter To Meter 3.0 mVDC Gas pressure measurement (0.1 resolution mode) 0.8 mVDC -0.03 inWC Static pressure across a blower (0.01 resolution mode) 0.8mVDC / 10 = 0.08inWC v13 RMA316 Manometer Adapter Used for 5/16" pressure outlet ports. 1. Shut off main gas supply to furnace. 2. Move the gas valve switch to "OFF" position. 3. Use a 3/32" hex wrench to loosen the outlet pressure tap screw. Rotate screw counterclockwise one revolution to open. 4. Connect the ADMN2 tubing to the smaller (3/16") end of the adapter and the other (5/16") end of the adapter into the of the adapter tube. 5. Slide the 5/16" adapter tube over the outlet pressure boss (port) to seal. Overlap the pressure boss by at least 3/8" to prevent leakage. Smaller end (3/16”) of the adapter connects to ADMN2 tubing. Remove the installed brass fitting from tube so the adapter can be inserted. Adapter (5/16” to 3/16”) Adapter Tube (5/16”) Outlet Pressure Tap Screw Inlet Pressure Tap Screw Outlet Pressure Boss Inlet Pressure Boss More Products From Fieldpiece CLAMP 300V CAT.III 400A AC Current Clamp Stick Meter This is the heart of modular expandability. In addition to being a full functioning multimeter, any accessory head can be used with it. ! 1AAC / 1mVAC 400AAC MAX ACH4 A & ir He Te Velo ad mp cit era y tu re ON LE AB ST n a tiot g lin er rig rhea bcoo 34A R1 R ef pe Su Su d an LO TT BA EN GL H IS SH En glish Ft /m M M in et PH ric ºF KM M/s OFF /h r ºC IC TR ME SC A 04 R4 SH SC p or m re Te su Pr es t Se M AT T/C l Ca RE AD ag se e c) 4 AA V3 AU OFFTO - T/C CAL AU TO OFF 2 AOX2 °F TEMP %CO %O 2 °C ON Lo Batt GAS LC 10 D X 0 BA TT Model HS36 Re tim al e OIL BLE Fue l Typ e STA ON LO is h Av (16 er ric Accessory heads are the sensors of multiple parameters measured by technicians every day. They plug into a mV range (depending on the head) of a multimeter. The multimeter will display whatever the head is measuring. Instead of having to purchase and carry a separate instrument for each parameter, a technician can use multiple heads and a single multimeter to do the job. TO AUOFF X2 AS Comb Chec ustio n k En gl M et Accessory Heads Modular Expandability Modular expandability is ability for accessory heads and meters to change configurations to match the various needs of an HVAC/R technician. Accessory heads (the sensors) send out a mV signal, which represents the value of the measurement, to whatever meter is attached to it. Heads can attach directly to the top of a Stick meter, DL3 data logger, or EHDL1. They can also plug into any meter with mV ranges using ASLS2 leads. Non contact voltage Magnetic hanger Autoranging Backlight Temperature Volts, amps, ohms Frequency Microfarads aaaaaaa aaaaa aaaaa a a a aaaa a aa aa a aa aaa aaaaa LCD X 100 READ ON Combustion Check STABLE Lo Batt GAS Fuel Type Metric English Metric Ft/min MPH ºF M/s KM/hr ºC OIL Real time TEMP %CO2 %O2 OFF AUTO OFF AUTOOFF AAV3 Includes: HS36 Meter ACH4 Current Clamp ATB1 K-type TCouple ADLS2 Deluxe Leads ANC1 Case ON LO BATT Air Velocity & Temperature Average English (16 sec) Head ARH4 °C °F AOX2 Here are four of the many heads available: AAV3 Air Velocity and Temperature ARH4 Digital Psychrometer AVG2 Digital Vacuum Gauge AOX2 Combustion Check T/C AUTO CAL OFF ASP2 Static Pressure Probe For accurate static pressure measurements. 1. Connect ASP2 to ADMN2 via hose. 2. Insert ASP2 into drilled or pre-existing 1/4” hole. 3. Use allignment arrow on the ASP2 to direct probe into the air stream. 4. Read static pressure measurements from “stickmeter”, EHDL1 or record with DL3. 5. Remove probe and patch hole.