Download LDP-3811 Programming Examples
Transcript
CHAPTER 3 COMMAND REFERENCE 3811 Device-Dependent Commands RANge? FRONT PANEL REMOTE The RANge? query returns the value of the laser drive current range. Parameters None. The response will be in the form: Notes Examples SECURE - where the response is an <nrf value>; 200 means 200 mA, and 500 means 500 mA range. The resolution and accuracy of the laser limit current are dependent on the laser drive current range. “RAN?” - response: 200, means the laser drive current output current range is 200 mA. “:Range?” - response: 500, means the laser drive current output current range is 500 mA. FRONT PANEL REMOTE The SECURE command allows the service technician access to the protected user data command, *PUD, which is a common command. This data is normally changed only at the factory, and therefore the SECURE command is not needed by the user. SET: FRONT PANEL REMOTE The SET: command path is used to get to the 3811’s laser set point queries. The following commands may be reached directly from the SET: command path. SET: CDC? SET: LDI? SET: PRI? 64 LDP-3811