Error Information General Error Messages 7 Compensation Sensor Open The ambient-compensation sensor (NTC) on the main board in the module has failed (open circuit). The resistance across the temperature compensation sensor (NTC) on the main board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated. Probable cause Suggested actions 1 Defective main board. Please contact your Agilent service representative. Compensation Sensor Short The ambient-compensation sensor (NTC) on the main board in the module has failed (short circuit). The resistance across the temperature compensation sensor (NTC) on the main board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated. Probable cause Suggested actions 1 Defective main board. Please contact your Agilent service representative. 1260 Nanoflow Pump User Manual 93