Looking more closer at the peak-table file and other reports is becomes clear
why Br-80 is not found. The highest peak of Br-80 is very close to a high peak
of Br-82 and is therefore not found.
In this figure (3.40) one of the energies is printed in a green font.
This means that the interference correction for this line can be checked. By
clicking at this figure with your mouse a special form will appear as shown in
figure 3.41. The standard interference correction is more or less qualitative
and meant for only nuclide identification. This option will allow a much better
correction and allows the user to check first the quality of the average
concentration of the interfering elements.
Figure 3.41: Quantitative interference correction.
You can accept the result or reset the old values. To make the best possible
result you should check the results of all involved nuclides. In this particular
case the correction is not really necessary and also not very good, so deselection of the bromine line (in figure 3.40) is a better and simpler action.
After the last element has been checked the screen given in figure 3.42 will
appear. Here the total spectrum countrate is given for each measurement as
Chapter 3: Analysis Evaluation