Download Kayzero for Windows Manual,first draft

Transcript
Explanation of some of the data in the KAYZERO Library
M/D (mother/daughter) relations and factor F1-F5
k0codes:
=======
1: recommended k0-factor
2: tentative (measured in only one laboratory)
3: theoretical k0-factor (for interference correction purposes only)
4: single escape peaks, with k0-factor of original peak (added according1)
5: double escape peaks, with k0-factor of original peak (added according1)
7: effective energy with recommended k0-factor
8: effective energy with non-recommended k0-factor
Eres :
=====
The mean resonance energy; if Eres is zero KAYZERO will use Q0-1 instead
of Q0(α).
Mother/Daughter relations:
=====================
The M/D relations are in the KAYZERO Library numbered from 1 to 19.
These number correspond to the roman numbers in the k0-standardization
literature of De Corte and Simonits2 . In Table A7.1 these numbers and the
factors F1-F5 given in the library are explained.
Half-lifes are given in minutes
======================
k0 factors/ k0codes point of attention
======================
Some k0 factors are only defined for an artificial effective gamma peak (in fact
a combination of two gamma lines close together). These gamma lines are
given in table A7.2. There are two types:
A: The lines can be separated using modern gamma-ray spectrometers and
deconvolution software. In this case the two separate k0 factors are given as
well as the k0 factor for the summation. In the latter case the k0 code is 7or 8.
B: The lines can not be separated.
Appendix 7: KAYZERO Library
-A26-