Download vDot-4 Manual

Transcript
Figure 6.2: Calibration Target
6.4 Measuring the Target
6.4.1 Using Genlin
If you do not have a strip-reading densitometer such as the
DTP-41 supported by Genlin, and you wish to use a manual
Status T densitometer, skip this section and go to 6.5, Entering
Measurement Data.
If you have the DTP-41 or similar instrument, use the Genlin
application installed with your RIP. Locate the Genlin icon
inside the RIP folder, launch it, and then select Configure from
the menu bar.
44
vDot4 Installation and Reference Manual