Download User Manual
Transcript
11. Instructions of rotor and tube 11.1.1 Rotor structure Knob Figure 11-1 the rotor structure Locking nut Seal ring Cover Shaft hole 11.1.2 Available rotors and adapters All rotors are used for bio-safe when the rotor lid was tightened with the rotor, centrifuge tubes will be enclosed into rotor to ensure the sample does not leak in centrifugal process. If rotor lid is not available, the rotor would be no bio-sealing function. The rotors can be used as follows: Rotor type 1 2 3 ID code Tube/bottle 2/1.5ml 0.2ml PCR 0.5ml micro 400.003.222 0.5ml micro PCR8 serial 400.003.223 PCR8 serial Table 11.1 Rotors and adapters 400.003.220 Adapter 400.003.250 400.003.251 400.003.252 Maximu m speed (rpm) 15000 15000 15000 15000 15000 15000 Maximum centrifugal radius rmax(cm) Maximum RCF RCF (×g) 8.5 6.9 7.6 8.5 7.6 6.5/7.2 21380 17350 19100 21380 19100 16350/18100 20