ANALYZER
QuickStart Guide
by
Study the User Manual
thoroughly before using
the QuickStart Guide
CUSTOMER SERVICE
If the Analyzer does not perform as expected, contact Quidel Technical Support (800) 874-1517
(in the U.S.), (858) 552-1100 (outside the U.S.) or your local distributor. [email protected]
UNPACK ANALYZER
◗ Inspect the shipping container for obvious shipping damage prior to opening.
◗ Unpack the shipping container and inspect the unit for damage.
Paper Cover
4AA Batteries
BATTERY Gold
BATTERY Gold
SUPER ALKALINE
LRG AA SIZE 1.5V
BATTERY Gold
SUPER ALKALINE
SUPER ALKALINE
LRG AA SIZE 1.5V
BATTERY Gold
Print
SUPER ALKALINE
ed
r Fe
LRG AA SIZE 1.5V
Pape
LRG AA SIZE 1.5V
Sofia Analyzer
by
Paper Roll
Eject
AC Power Cord
DC Power Cord
UNPACK INSTALLATION PACK
◗ Inspect the shipping container for obvious shipping damage prior to opening.
◗ Unpack the shipping container and inspect the unit for damage.
Barcode Scanner
Barcode Scanner Sta
Stand
and
CD
quidel.com/fluorescence
Literature Translations
1205700 (09/11)
Calibration
Cassette
Pouch
SD
D Card
Calibration
on
Cassette
CONNECT TO POWER AND TURN ON
Place the Sofia Analyzer on the counter top within reach of an electrical outlet.
The unit is portable and can be moved to a suitable location for testing.
1
Plug the DC power cord into
the power port in the back
of the Analyzer.
2
Plug the appropriate, country specific
AC power cord into the DC Power Cord
and appropriate electrical outlet.
3
Plug the Barcode scanner
into the PS2 port in the back
of the Analyzer. Note: The
Barcode scanner may have
2 plugs; one will remain
unplugged.
4
Turn the Analyzer on using the power
switch located on the rear panel. Hold
the toggle switch down until the
screen illuminates (1–2 seconds).
Power Switch
CALIBRATION CHECK
◗ Use the up/down keys to highlight CALIBRATION from the menu.
◗ Press the soft key under the lower right corner of the display screen under SELECT.
◗ Remove Calibration Cassette from the pouch.
y
06-01-2011 | 14:46
Main Menu
Run Test
B
ad
e
FL r
A+ U
Re
Run QC
Review Data
Calibration
Paper Feed
Supervisor Menu
Change Mode
Select
Eject
Insert the Calibration Cassette into the
Analyzer and close the drawer. The Analyzer
will perform the calibration automatically.
Print
06-01-2011 | 14:46
Start Calibration
A+B
Reader
FLU
Please insert cassette and close drawer.
Cancel
The Analyzer display will indicate when the
calibration is completed.
NOTE: If calibration cannot be completed
successfully, notify the on-site
Supervisor or contact Quidel Service for
assistance.
06-01-2011 | 14:46
Calibration Results
Calibration passed.
After successful calibration, select OK to
return to the Main Menu.
Press eject to remove and return the
Calibration Cassette to the storage pouch.
Important: Store the Calibration
Cassette in the provided storage pouch
between uses.
Run QC according to assay specific package insert.
OK
User ID
Patient ID
Order #
User ID
Run QC
Review
Patient
Results
Review
QC
Results
Review
Calibration
Results
Review
Data
View
Lot
Status
View
Test
Types
User ID
Calibration
See
Supervisor
Structure
Supervisor
Menu
Change
Mode
User Menu Structure
Run Test
Main Menu
MENU STRUCTURES
Quidel Corporation
San Diego, CA 92121 USA
Set Language
& Sound
Set Location
Name
Save Logs
to SD Card
Load/
Save
Set Date
Set Calibration
Reminder
Delete User
Save Data
Set Entry
Requirements &
Timeout
Load Data
Edit User
Soft Key:
Options
Manage
Users
Set Default Mode
& Print Option
Add User
Load Test
Type
Set Time
Set User Access
Rights
Set Network
Configuration
View Error Log
Statistics
General
Settings
View User
History
Load Language
File
Update
View Info
Load Analyzer
Software
Supervisor
Menu
Set LIS Address
Settings
Supervisor Menu Structure
1201100EN00 (09/11)