Download Spinmaster FFC2000 V1073
Transcript
Spinmaster FFC-2000 - Reference manual page 3 of 3 1. Spinmaster console • • • • Variable Temperature Control (VTC) unit for sample temperature control Pulse programmer and signal acquisition (AQM) unit RF unit: - DDS RF synthesizer - RF Receiver - RF Power transmitter Power supply unit 2. Magnet Power Supply • • • • • • Power MOSFET Banks Reference and Field Regulation units Interlocks system Current and Magnet temperature display Magnet temperature compensation system Electronic power supply for all the above-mentioned systems. 3. Cooling System • • • Magnet liquid cooling pump Cooling liquid tank reservoir Cooling liquids, heat exchanger 4. Magnet and Probe • • • Magnet Probe and air flux heater Preamplifier and front end box 5. Computer and AcqNMR32 software for data acquisition. Specifications of the Spinmaster - FFC NMR Relaxometer Magnet Maximum field Homogeneity : : : Power Supply Max. Power Max. Current Switchable levels : : : : Switching time Mains : : Cooling Unit Water request : : Copyright © Stelar s.r.l 2001 Special design, aluminium air coil system 0.5 Tesla better than 100 ppm over 1cm3 15 kW 350 A 4 levels (off, Bpol, Brelax, Bacq) each level is software controlled with 12 bit resolution < 0.15 ms / MHz 400 VAC / 22 kW and 220 VAC / 2 kW Dual independent, thermally coupled loops. Secondary is a closed circuit using a cooling fluid at room temperature. Primary circuit uses tap water with regulated consumption. 2000 l/hr of water at 20°C (0.7bar) at maximum power