Manual Cleaning
ATTACHMENTS
Caution: Manual Cleaning Instructions for all
Attachments except the Minimal Access (MA)
Attachments and bearing sleeves.
Note: “DO NOT IMMERSE” as marked on attachments
is for OR personnel only.
Detergents
• Alkaline detergents with a maximum pH of 11 are
recommended for the manual washing of the attachments.
• The process, detergent and any additives must be discussed
with detergent manufacturer.
Please refer to cleaning instructions provided with the
MA attachments for Minimal Access (MA) Attachments,
Minimal Access Drivers (MA-D20, MA-DRIVER), Minimal
Access Bearing Sleeves (MA-10S, MA-10C, MA-15S,
MA-15C, MA-15ST, & MA-19ST).
Warning: Wear eye protection and any additional
PPE required.
Caution: Do not use alkaline detergent with a pH
greater than 11. Alkaline detergents attack grease
and seals which can increase wear and cause device
malfunction. Please note use of alkaline detergents can
cause color to fade but this does not impair function of
device. Do not use ultrasonic equipment or corrosive
or harsh chemical soaps. Do not rinse attachment with
saline solution.
1
Preparation
Assemble all necessary supplies including Attachment
Cleaning Brush (ACB), sponge, lint-free cloth and soft bristle
brush. Remove the attachment and dissection tool from the
handpiece (Figure 1).
CSR60 only: The red storage cap provided with the
Perforator Driver is to prevent damage to the shaft.
The cap is removed for cleaning, washing and sterilization
(Figure 2). The cap is used for storage purposes only.
Figure 1
Figure 2
2
Immerse
Fully immerse the attachment in an alkaline cleaner
(non-chlorinated detergent), prepared as described on the
product label, at room temperature, in a suitable container
and agitate for 15 seconds (Figure 3).
Figure 3
Anspach eMax 2 Plus System
User’s Manual
DePuy Synthes Power Tools
37