Download Alternative Design 2 - Biomedical Engineering
Transcript
Figure 13. Pulsatile signals found in the intensity of detected light [15]. One pulse is one cardiac cycle. The microprocessor will count the pulses to determine heart rate (beats per minute), which will be displayed on an LCD screen. 1.2.3 Non Invasive Blood Pressure Blood pressure will be automatically measured through the oscillometric method [14]. This is done by wrapping a blood pressure cuff around the upper arm and inflating it until the pressure around the arm due to the cuff collapses (or occludes) the brachial artery. The cuff is then slowly deflated. As the cuff deflates, blood starts pumping through the brachial artery causing minute vibrations of .5 to 1 mmHg in the cuff [2]. The pressure at which these vibrations start is the systolic pressure, and the pressure at which they stop is the diastolic pressure [3]. The block diagram in Fig. 16 illustrates how this method will be used to measure blood pressure in the accessible vital signs monitoring system. Each system in the flow chart is described in more detail in the following paragraphs. When the blood pressure “Start” button on the vital signs monitor is pressed, the blood pressure cuff will be inflated to about 40mmHg above normal (160mmHg). The blood pressure cuff used will be a DRE Adult single lumen cuff from DREMed.com at a cost of $37.00, before shipping and tax (Fig. 14). http://www.dremed.com/catalog/ product_info.php/cPath/56_121_241_242/products_id/194 Figure 14. DRE Adult Single Lumen Blood Pressure Cuff 13