Download Manual instruction - WLX, WTX balances
Transcript
Or using the parameter: settings/ RS-232/Aut. printout/continuous Frame format: 1 2 3 4 5 6 7-15 16 S I space stability space sign mass space Stability character Sign Mass Unit Command 17 18 unit 19 20 21 CR LF [space] if stable [?] if not stable [^] if an indication over the range [v] if an indication below the range [space] for positive values [-] for negative values 9 characters justified to the right 3 characters justified to the left 3 characters justified to the left Notice: The frame length is constant (21 characters). 22.4.3. Continuous transmission „with pauses” Continuous transmission „with pauses” allows users to determine the time interval between the subsequent frames in the parameter interval. His type of printouts can be enabled only in parameters: Settings/RS-232/ Aut. printouts/with interv. The frame format is the same as in basic simple transmission. The interval is set in: Setting/RS-232/Interval 22.4.4. Stable result transmission The scale automatically generates after each stable result separated by a stable result below the Min. mass that can be set in parameters. The intended use of this printout is for slip printers so it doesn’t have a starting characters and the length can vary from 18 to 20 characters. Standard length is 18 characters, but if the parameter digit marking has been enabled, the length increases to 20 characters, because additional brackets are printed e.g.: - 58 -