7.2.2.4 Post-Treatment of Sterilization Material When unloaded from the sterilizer, all material sterilized with steam contains residual moisture, which should evaporate freely when the material is cooled. Compared to containers made from aluminium or wire baskets, the conditions for stainless steel and chromium plated brass are particularly unfavourable. Overloading the sterilizer has negative consequences for the drying result. Passing the commended times of sterilization without additional cooling the drying of the sterilization goods is significantly reduced so that additional drying times (e. g. 510 minutes evaporation with slightly opened door) are recommended (see chapter 6). INDICATION: If dishes or trays are used, they should not be set onto flat and solid surfaces. The condensate coat created by cooling down to room temperature may subsequently cause soaking of the wrapped sterilization material. Therefore, dishes and trays are better set on grids or other structured surfaces. CAUTION! When loading or unloading the chamber, the door plate and the chamber walls are preheated. Do not touch these areas! Danger of burning! A35SE1-2 Vers. 02 / 04.98 WEBECO Hygiene in Medizin und Labor GmbH und Co. KG 31