OBJ_BUCH-11-001.book Page 22 Tuesday, May 25, 2004 10:30 AM Phyaction U 4.5.3 1. 2. 3. 4. 5. 6. 7. Read-out values for ultrasound therapy Channel Ultrasound therapy Type of US head Remaining treatment time Îset Ppk Contact of the US head 1 2 3 4 5 6 7 Contact of the US head The contact of the US head with the skin: • : Bad contact, US head switched off (0 W). • : Bad contact. • : Sufficient contact. • : Good contact. • : Very good contact Test the US head if its conduction is bad. See §5.1.1. Îset (W/cm2) The power (W) of the US head per cm2. Ppk (W) The peak power of the US head (Îset * ERA). The peak power delivered therefore depends on the size of the US head and the contact with the skin. This value is 0.0 W if the contact with the skin is bad. In this case, the ultrasound treatment of the equipment is stopped to prevent overheating of the transducer. 4.5.4 Parameters for ultrasound therapy Treatment time (mm:ss) The duration of the treatment. Duty cycle (10, 20, 30, 40, 50%, continuous) Ratio of the pulse duration to the period duration. • Continuous: Continuous ultrasound (100%). • 10, 20, 30, 40, 50%: Pulsating ultrasound. Select a high duty cycle for an intensive treatment. Select a low duty cycle for a mild treatment. 22