PSM3750 user manual 4.3 Zero compensation There are 3 levels of zero compensation: Trim out the dc offset in the input amplifier chain. Measure any remaining offset and compensate. Measure parasitic external values and compensate. The trim of the dc offset in the input amplifier chain can be manually invoked at any time with the ZERO key, or over the RS232 with the REZERO command. The measurement of the remaining offset also happens when the offset is trimmed but is also repeated at regular intervals when using a measurement function that requires dc accuracy (such as the rms voltmeter). This is to compensate for any thermal drift in the amplifier chain. This repeated autozero function can be disabled via the SYSTEM OPTIONS menu. The compensation for parasitic external values (for example to compensate for the capacitance of the test leads when measuring capacitance) is invoked manually by the ZERO key. Refer to each function section for the function specific operations. Any compensation values are stored along with the instrument configuration when a program is stored. To restore operation without function specific compensation press ZERO then select the clear option. 4-3