Download Model 48i - User Manual
Transcript
Operation Diagnostics Menu PRESSURE: RANGE Flow AVG DIAGS ALARM The Flow screen (read only) displays the flow rate. The flow is measured by an internal flow sensor. For more information, see Chapter 1, “Introduction”. ● In the Main Menu, choose Diagnostics > Flow. FLOW: RANGE Sample/Reference Ratio 753.0 mmHg 1.108 L/min AVG DIAGS ALARM The Sample/Reference Ratio screen (read only) displays the ratio of the intensities of the light source through the sample wavelength and reference wavelength of the bandpass filter wheel. Normally, when zero air is being sampled, the sample/reference ratio is between 1.14 and 1.18. A ratio outside may indicate that the filter wheel is dirty or the infrared source is degraded. ● In the Main Menu, choose Diagnostics > Sample/Ref Ratio. SAMPLE REFERENCE RATIO: S/R: RANGE AGC Intensity 3-64 Model 48i Instruction Manual 1.161111 AVG DIAGS ALARM The AGC Intensity screen (read only) displays the intensity (in Hertz) of the reference channel Automatic Gain Control (AGC) circuit. The AGC circuit optimizes the noise and resolution levels of the Model 48i. The AGC intensity reading should be about 250,000 Hertz. Thermo Fisher Scientific