Chapter 4 for a few hours. It is necessary to know the total volume of the closed-loop system. For this purpose, the volume inside the RAD7 may be taken as 1L. The initial rate of increase in radon concentration (neglecting the first 15-min cycle), multiplied by the volume, then gives the rate of emission of radon. A reduction in the slope, as the radon level builds up, may be due to leaks in the system, or to a reduction in the net emission. the thoron that gets fed back to the enclosure. Note that, for thoron, both the closed loop and open loop modes are steady-state measurements. With the open loop configuration, a steady, and known, flow rate must be established. If a cylinder of ‘zero’ air, or nitrogen, is used, then the RAD7 internal pump may be set to Off, and the pressure reduction valve used to control the flow rate. With ambient air, the internal pump may be set to On, for a continuos flow. In both cases, a flow meter is required. The alternative procedures are similar to those for bulk emission measurement, as described above. Once the total emission rate, within the enclosure, has been calculated, it may be divided by the area of the surface within the sealed boundary, to determine the emission per unit area. 4.7.4 For these measurements, an enclosure, with two hose connectors, has one side open, which is sealed against the surface under investigation. A steady state should be achieved and a long-term measurement may be made. The rate of emission will equal the radon concentration times the flow rate. The precision will depend on the concentration and the duration of the measurement. 4.7.3 Surface Emission 4.8 Thoron Measurement 4.8.1 Thoron and Radon Thoron is an isotope of the element radon having an atomic mass of 220, so it is also known as radon-220. The word "radon" without a mass number almost always refers to radon-222. Thoron and radon have very similar properties. They are both chemically inert radioactive gases that occur naturally from the decay of radioactive elements in soils and minerals. Both thoron and radon are members of decay chains, or long sequences of radioactive decay. Bulk Emissions An airtight box with two hose connectors, well separated, will usually work fine. The material is placed in the box. The box is connected, by tubing from one connector, to the laboratory drying unit, and thence to the inlet filter on the RAD7. The other box connector has tubing attached, which is either connected to the RAD7 outlet, for closed-loop operation, or, for open- While radon results from the decay of natural uranium, thoron results from the decay of natural thorium. Both uranium and thorium are commonly found in soils and minerals, sometimes separately, sometimes together. The radioactive gases radon and thoron that are produced in these soils and minerals can diffuse out of the material and travel long distances before they themselves decay. Both radon and thoron decay into radioactive decay products, or progeny, of polonium, lead, and bismuth before finally reaching stable forms as lead. loop operation, connected to a cylinder of zero gas, or left open to ambient air. Note that bulk emissions are affected by pressure fluctuations and by temperature and humidity. All these parameters can, and should, be controlled in both the closed-loop and open-loop configurations. Radon emission is also dependent on the grain size of loose materials, and porosity of any bulk material. Thoron and radon and their respective progeny differ very significantly in their half-lives and in the energies of their radiations. While radon has a half-life of nearly 4 days, thoron has a half-life of only 55 seconds. Since thoron is so short lived, it cannot travel as far from its source as radon can before it decays. It is commonly observed that Thoron emission can also be measured in same apparatus. In the open loop mode, a correction will be required for the decay of the thoron during the time between emission and measurement in the RAD7. In the closed loop mode, another correction should be made for the proportion of 53