7.4.9 Application –> Identification
ID1, ID2, ID3
Labelling the identification soft keys
Soft key description
Entering a soft key label for the identification soft keys with max.
5 characters.
E.g., soft keys "User, "Art.No.", "Lot" instead of ID1, ID2, ID3
Printout description
Entering a description of the identification for printout with max.
40 characters.
E.g., "User name", "Article number", "Lot number" on the printout instead
of ID1, ID2, ID3
7.4.10 Application –> Statistics
Statistics
Selecting statistical information to be displayed or printed
Standard dev
Standard deviation of a weighing series
Std. dev. good
Standard deviation of all samples within tolerances of a weighing series
Mean value
Mean value of a weighing series
Mean value good
Mean value of all samples within tolerances of a weighing series
Max. value
Maximum weight value of a weighing series
Min. value
Minimum weight value of a weighing series
Median
Weight value separating the higher half of a weighing series from the
lower half
% ratio per class
Percentage of good, high and low items in a weighing series
# per class
Number of good, high or low items in a weighing series
Note
In the factory setting all items are activated