Download User`s Manual
Transcript
User’s Manual AXFA14G/C Magnetic Flowmeter Remote Converter [Hardware Edition/Software Edition] AXF Magnetic Flowmeter Integral Flowmeter [Software Edition] Manual Change No. 06-015-2E Please use the attached sheets for the pages listed below in IM 01E20C02-01E (7 th). Page and Item Page 5-2 5.2 Display Unit Setting Methods Contents of Correction Corrected the flow chart. Changed the note of entry mode. Dec. 4, 2006 Yokogawa Electric Corporation 5. BASIC OPERATING PROCEDURES 5.2 Display Unit Setting Methods NOTE Before changing any settings, be sure to check the corresponding setting details in Chapter 6: Parameter Description. 5.2.1 Display Mode → Setting Mode Display Mode will be adopted when the power is turned on, and the Setting Mode can be activated using the following procedure. Parameter Search Mode Select “Yes” (2 seconds or more) Display Mode Major Item Parameter Search Mode Entry Mode Parameter Replacement Mode Sub-item Parameter Search Mode + + Select “No” Setting modes * The term "mode" is used to refer to a situation where display and setting are possible. + Press the SET switch while holding down the SHIFT switch. F0502.EPS Sample Display: Procedure for moving from Display Mode to Setting Mode 1st line: Actual instantaneous flow rate [FR] Start: Display Mode 100.000 0 3274m3 2nd line: Bar graph indicating instantaneous flow rate FR m3/h 100 FTL 3rd line: Totalized forward flow rate [FTL] Setting switches Hold the Touch (*1) switch for 2 seconds. and switches simultaneously. Entry Mode Touch SHIFT + key NOTE (*1) The Major Item Parameter which is set just before will be shown when entrying the Setting Mode again within 1 minute after returning from Setting Mode to Display Mode. F0503-1.EPS 5-2 IM 01E20C02-01E