EN 61010-1 + EN 61010-2-101
Clause
Requirement + Test
Result - Remark
Impact energy level and corresponding IK code:
No IK coded
Verdict
N/A
Non-metallic ENCLOSURES cooled to minimum RATED
ambient temperature if below 2 °C
P
8.3
Drop test
P
8.3.1
Other than HAND-HELD and DIRECT-PLUG-IN EQUIPMENT
Floor-standing equipment
Tests conducted with a drop height or angle of ..... :
Drop from 30
HAND-HELD and DIRECT-PLUG-IN EQUIPMENT
Floor-standing equipment
8.3.2
8.101
9
9.1
o
P
N/A
Non-metallic ENCLOSURES cooled to minimum RATED
ambient temperature if below 2 °C
N/A
Drop test conducted with an height of 1 m
N/A
Transport and storage (EN 61010-2-101)
P
When packed in the manufacturer’s packaging,
equipment shall not cause a HAZARD during
NORMAL USE after transport or storage in the
conditions specified by the manufacturer (see 5.1.101
and 5.4.101). (EN 61010-2-101)
P
PROTECTION AGAINST THE SPREAD OF FIRE
P
No spread of fire in NORMAL and SINGLE FAULT
CONDITION
MAINS supplied equipment meets requirements of 9.6
additionally
There is no spread of fire
outside the equipment in
normal or single fault
condition
P
OCP is provided in the
adapter
P
P
Conformity is checked by minimum one or a
combination of the following (see Figure 11):
a) Fault test of 4.4; or
9.2
P
(see Form A.1 and Form
A.2)
P
b) Application of 9.2 (eliminating or reducing the
sources of ignition); or
P
c) Application of 9.2 (containment of fire within the
equipment)
P
Eliminating or reducing the sources of ignition within the Considered to be reduced
equipment
to a tolerable level
P
a) 1) Limited-energy circuit (see 9.4); or
2) BASIC INSULATION provided for parts of
different potential; or
Not determine to Limitedenergy circuit
N/A
(see Form A.5 and A.14)
Insulation between parts at
different potentials meets
the requirements for basic
insulation
P