Download PowerMon

Transcript
30
4.2
Bran+Luebbe PowerMon
Error messages overview
Query
System
Error
class
1
2
Liquid
detector
2
1
Heating
1
2
Error message
Memory error
CAN Bus error
Calculation formula 1 missing
Calculation formula 2 missing
Conductivity sensor failed
Pump speed 1 incorrect
Leak alarm
Photometer failure
Measuring chain
UV-Digestor failed
Consequence
STOP – Error message – System
alarm
Error message – device continues
operation
Measurement stops – Error message – selection of next MP.
Sample 1 - 6 lack alarm
Dilution lack alarm
Calibration solution lack alarm Calibration program stops
- Error message
- Measurement continues with old
calibration values.
Reagent 1 - 5 lack alarm
STOP – Error message – System
alarm
Temperature sensor shortSTOP - Error message – System
circuit
alarm
Temperature sensor cable
broken
Heating safety shutdown
Temperature too high
Error message – device continues
operation
Temperature too low
Operating
hours
counter
2
Warning limit per output*)
1 or 2*)
Operation limit per output*)
Limits
2
MP 1 – 6 upper limit exceeded
MP 1 – 5 lower limit exceeded
Upper warning limit
Error message - Calibration will be
repeated once.
Lower warning limit
If the error occurs again – Error
message – Measurement continues
Upper operation limit
Error message - Calibration will be
repeated once.
Lower operation limit
If the error occurs again – Error
message – System alarm
Calibration 2
1
Error message – device continues
operation
STOP – Error message – System
alarm or Error message – device
continues operation, *): defined by
operator
Error message – device continues
operation