CE318-T Photometer User Operation Manual 4. Loosen the 2 nuts maintaining the probe clamp 5. Withdraw the screw and the wet sensor with is wire from the protection case 6. Slide the wet sensor to get it back and replace it. 5.2.2. Robot leveling adjustment After a PARK scenario, even if the spirit level is well-centered, it might happen that the V-shape part of the robot is not perfectly horizontal. You can check with a leveler. If the case happens: 1. Loosen the 2 hexagon socket screws. 2. Rotate the robot horizontal axis so that the V-shaped part is perfectly horizontal. Check with the leveler. 3. Tighten the screws and check is the v-shape part is perfectly horizontal after having launched a PARK scenario. Revision 3.2 April 2015 46/65