Download "user manual"

Transcript
Ch.6 Web Maintenance Mode
Item
FORMAT
Default
Descriptions
CH01 : CENTER
This setting determines space allocation between frames in multi-
CH02 : EVEN
format printing.
CH03 : CENTER
CH04~16 : EVEN
"CENTER"
"EVEN"
: Central allocation.
: Even allocation.
"H CENTER"
: Central allocation (When diagnostic device is of
Hitachi manufacture)
"H EVEN"
: Even allocation (When diagnostic device is of
Hitachi manufacture)
The methods of allocation depending on format type are shown below.
<Central Allocation>
Spaces between frames are equal with margins at left and right
sides expanded.
<Even Allocation>
Spaces between frames are expanded, leaving the
upper/lower/left/right margins unchanged.
Item
DEFAULT DMIN/DMAX
Default
CH01~16 : DICOM
Descriptions
Sets acquisition of maximum/minimum density values used only for
image processing.
"DICOM"
: The value specified by the diagnostic device is
"P-CON"
: Maximum/minimum density data included in LUT
used.
settings is used. The LUT used is specified from
the diagnostic device.
DICOM PRIORITY
CH01~16 : OFF
Sets acquisition of parameters used for printing.
"ON"
: Uses values input in print condition settings.
"OFF"
: Uses values input in the diagnostic device.
The following items may be set.
- LUT
- SMOOTHING
- DENSITY MAX
- DENSITY MIN
- POLARITY (NEGA/POSI)
- ORIENTATE
- TRIM
- BORDER (Background density)
- SORTER (Sorter destination bin)
< 186 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11