Download DiCromat II User`s Manual

Transcript
IX. Appendix
Quick Start Card Facsimile
DiCromat II
CALIBRATION MODE
THIS MODE CAN BE ENTERED IN TWO WAYS:
• FROM INITIALIZATION MODE: Answer
“No” when asked “Go to Run mode?”
• FROM RUN MODE: Press “Go to Cal.”
1. Enter the desired Set Point #.
2. Confirm or enter the following:
•
•
•
3.
Name of the sample to test.
Cell factor of the probe - This number is
imprinted on the probe.
Dilution Factor
Select unit of measurement with “YES.” “NO” will scroll through
alternatives.
a. Percent Salt (A)
b. Degrees Salometer (S)
c. Conductivity (K)
d. Molar Concentration (M)
e. Concentration (Cs)
4.
Confirm / Enter full scale output value. (10V OUT). Enter “0” if not in
use.
• This is important only when the analog output is being utilized.
5.
Sample Calibration
• Test known sample by immersing dip-in probe or using pour-through
probe. (READING WILL FLASH UNTIL CONFIRMED)
• If reading is different from known value, answer “NO.”
• Enter actual titrated value.
• Confirm “OK” by entering “Yes.”
6.
Calibrate additional Set Points with “YES” answer. Answer “NO” to go to
RUN MODE.
The Noramar Company.
16