Download - Bio-Rad
Transcript
TROUBLESHOOTING SYSTEM INSTALLATION AND SETUP Problem Possible Cause Negative peaks. The sample is applied to the column in a different buffer from that used to equilibrate and elute the column. Refractive index changes may be responsible for the negative peaks. 1. Apply the sample in the same buffer used to equilibrate the column. The elution buffer may have a higher UV absorbance than the sample components. 2. Check the elution buffer’s UV absorbance. The injection valve, injection port, and loop were not rinsed out between sample injections. 1. Rinse valve, port, and loop with buffer. For example, if the sample loop contains 100 µl, then program a load/inject volume of 200 µl to completely flush the valve and loop. False or ghost peaks. Suggested Solutions 2. Always clean large volume dynamic loops, such as the Bio-Rad DynaLoop™, between use to avoid cross contamination. Strongly-retained sample 3. Use stringent elution conditions or consult the components may still be column’s user manual for an appropriate wash eluting from a previous step to remove such components between runs. run. 6-4