As dosing is a kind of dynamic state, which results in continuous changes in measurements, the averaged number of samples in the filter have an effect on the result. An example situation is illustrated below: The upper blue line represents results for n=1 samples in the filter buffer (averaging is off). The lower red line represents the same process when the filter is set to n=10. The difference depends on the dynamics (dosing rate) of changes either. The theory shows that the best filter parameter is n=1 because the setpoints can be compared with the current dosed mass. But in practice, there is a noise from different vibration sources registered and sometimes external forces connected with kinetic and potential energy of the poured material. It causes that the filter setting should be matched experimentally. Caution: 1. The subsequent readouts of the measured value from the A/D converter is performed every 100 ms. 2. This filter operates only in dosing procedures. - 29 -