Download WCA230A & WCA280A Programmer Manual

Transcript
:SYSTem Commands
:SYSTem:ERRor:CODE:ALL? (Query Only)
Returns all the unread error/event codes from the error/event queue, and removes
all the information from the queue. For details of the error messages, refer to
page 3–17.
Syntax
Arguments
Returns
:SYSTem:ERRor:CODE:ALL?
None
<ecode>{,<ecode>}
Where
<ecode>::=<NR1> is the error/event code, ranging from –32768 to 32767.
Measurement Modes
Examples
All
:SYSTem:ERRor:CODE:ALL?
might return -101,-108 of the error codes.
:SYSTem:ERRor:CODE[:NEXT]? (Query Only)
Returns the most recent unread error/event code from the error/event queue, and
removes that information from the queue. For details of the error messages, refer
to page 3–17.
Syntax
Arguments
Returns
Measurement Modes
Examples
2-532
:SYSTem:ERRor:CODE[:NEXT]?
None
<ecode>::=<NR1> is the error/event code, ranging from –32768 to 32767.
All
:SYSTem:ERRor:CODE:NEXT?
might return -101 of the error code.
WCA230A & WCA280A Programmer Manual