Download User`s Manual
Transcript
Performance Tests 4-5 PERFORMANCE TESTS INTRODUCTION The procedures in this section test the performance of the DS335. The first set tests the basic functionality of the DS335 from the front panel. The second set of tests actually measures the DS335's specifications. The results of each test may be recorded on the test sheet at the end of this section. NECESSARY EQUIPMENT The following equipment is necessary to complete the tests. The suggested equipment or its equivalent may be used. Instrument Critical Specifications Recommended Model Analog Oscilloscope 350 MHz Bandwidth Tektronix 2465 Time Interval Counter Frequency Range: 20 MHz minimum Time Interval Accuracy: 1ns minimum SRS SR620 FFT Spectrum Analyzer Frequency Range: DC to 100 kHz Amplitude Accuracy: ±0.2 dB Distortion: < 75 dB below reference SRS SR760 RF Spectrum Analyzer Frequency Range: 1 kHz to 100 MHz Amplitude: ±0.5 dB Distortion and Spurious: < -70 dB Anritsu MS2601/ HP4195A DC/AC Voltmeter 5 1/2 Digit DC accuracy True RMS AC to 100 kHz Fluke 8840A Thermal Converter Input Impedance: 50Ω Input Voltage: 3 Vrms Frequency: DC to 10 MHz Accuracy: ±0.05dB Ballantine 1395A-3 10 MHz Frequency Standard Frequency: 10 MHz ± .001 ppm Phase Noise: < -130 dBc @ 100Hz SRS FS700 50Ω Terminator 50Ω ± 0.2 %, 1 Watt HP 11048C Doubly Balanced Mixer Impedance: 50Ω Frequency: 1 - 20 MHz Mini-Circuits ZAD-3SH DS335 Synthesized Function Generator