Download photodetector module DM0089C data sheet ENS- ECH

Transcript
DM0089C data sheet
page 2
10 dynamic range
12 outline drawing (mm)
Typical counting sensitivity for DM0089C at 400 nm. Note
that automatic dead time correction has been applied at high
count rates. The dynamic range can be further extended at
low count rates by dark count subtraction in the software.
10
output counts s-1
10
10
8
10
5
6
3
1
8
13.5
4
Hirose HR10-7R-6P
6 way fixed plug
pin 1 USB 0 V
2 USB power
3 USB DM
4 USB DP
5 user out
6 trigger in
2
7
15.0 crs
6
Pin designations
10
10
10
5
power input
+4.75 V to +5.25 V
2.1 mm dc power
connector (centre
positive)
4
3
10
2
10
count
subtraction
darkdark
count
subtraction
1
10
-17
10
-16
10
-15
10
-14
10
-13
10
-12
10
-11
10
-10
10
4 holes,
M3 x 6.5 deep
on 25 centres
-9
incident light power W at 400 nm and 20 °C
±0.25
11 functional diagram
80.0
photomultiplier
photocathode
anode
dynodes
±0.25
15.7
+ve HV supply
disable
HV
+5V (red)
22.0 dia active area
amplifier
/
discriminator
0V (black)
trigger in (grey)
USB
type A plug
USB
interface
microcontroller
user out (violet)
±0.25
±0.25
32.0
56.5
13 installation and operation
counter
Each module is supplied with test data. Wherever possible
carry out installation in subdued light.
Remove the protective cap from the window of the module. If
necessary, clean the photomultiplier window using a lens
tissue moistened with alcohol. Do not use any other solvent.
Mount the module and make power input and signal
connections.
Do not expose the photomultiplier tube photocathode to
strong lights while the module is energised.
14 warning
Do not attempt to repair or dismantle this product. High voltage
used within the module presents an electric shock hazard.
Do not operate beyond the maximum ratings, or reverse the
input voltage; this may result in loss of performance or
permanent damage to the product.
Sens-Tech Limited
6A Langley Business Centre,
Station Road, Langley
Berkshire, SL3 8DS, UK
tel:
+44 (0)1753 214714
fax:
+44 (0)1753 214715
e-mail: [email protected]
The company reserves the right to modify these designs and
specifications without notice. Developmental devices are intended for
evaluation and no obligation is assumed for future manufacture.
While every effort is made to ensure accuracy of published information
the company cannot be held responsible for errors or consequences
arising therefrom.
an ISO 9001 registered company
www.sens-tech.com
SENS-TECH
SENSOR TECHNOLOGIES
© Sens-Tech Limited
DS_DM0089C Issue 3