Download Click here for manual
Transcript
Problems CAL and CO2 calibration are not displayed after connecting the CO2 sensor to the monitor. After the CO2 sensor is inserted, CO2 waveforms stay a flat line. Possible Causes The CO2 sensor is malfunctioning. Recommended Actions The CO2 sensor needs to be returned to manufacturer for service. The CO2 sensor is malfunctioning or not calibrated properly. The CO2 function is not working properly. The waveforms are erratic. The airway adapter is dirty inside or damaged. The airway adapter can not be cleaned. Re-calibrate the CO2 sensor, and then test it on yourself by breathing through it. If still the same, the CO2 sensor needs to be serviced. Insert another airway adapter, and then recalibrate the CO2 sensor. The airway adapter is semidisposable, and needs to be replaced frequently. 55