Transcript
Temperature Calibrator Specifications Millivolt Measurement and Source* Range Resolution -10 mV to 75 mV R Accuracy ±(0.025 % + 1 count) 0.01 mV S Maximum input voltage: 30 V Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C *Select this function by pressing T. The signal is available at the thermocouple miniplug connector. Temperature, Thermocouples Type J K T E Range B L Measure and Source Accuracies (ITS-90) -200 to 0 °C 1.0 °C 0 to 1200 °C 0.7 °C -200 to 0 °C 1.2 °C 0 to 1370 °C 0.8 °C -200 to 0 °C 1.2 °C 0 to 400 °C 0.8 °C -200 to 0 °C 0.9 °C 0 to 950 °C 0.7 °C U N -20 to 0 °C 2.5 °C 0 to 500 °C 1.8 °C 500 to 1750 °C 1.4 °C -20 to 0 °C 2.5 °C 0 to 500 °C 1.8 °C 500 to 1750 °C 1.5 °C 600 to 800 °C 2.2 °C 800 to 1000 °C 1.8 °C 1000 to 1800 °C 1.4 °C -200 to 0 °C 0.85 °C 0 to 900 °C 0.7 °C -200 to 0 °C 1.1 °C 0 to 400 °C 0.75 °C -200 to 0 °C 1.5 °C 0 to 1300 °C 0.9 °C Resolution: J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F 41