Download Explorer XP - Spectra
Transcript
Explorer XP Actively Q-Switched, High-Repetition Laser Systems READ:QSW:CNTS? This command reads the pulse counter in megacounts. Unit: Mcnts Examples: READ:QSW.CNTS?<CR> 34.4Mcnts<CR><LF> How many pulses have there been since the last pulse count reset? The system has issued 34,400,000 pulses. READ:SHG:TEMP? READ:THG:TEMP? These queries retun the measured SHG or THG temperature in counts. Unit: cnts (counts) Example: READ:SHG:TEMP?<CR> 1649cnts<CR><LF> What is the actual SHG temperature? The actual SGH temperature is 1649 counts. STAT:APORT:SIGNAL? This query returns the status of the ANALOG I/O interface pins. Use this query to find the activity level of the individual signals applied to the ANALOG I/O port (refer to Table B-5). Individual bit position = 1: signal on pin is active Individual bit position = 0: signal on pin is not active Example: STAT:APORT:SIGNAL?<CR> 64<CR><LF> What is the status of the lines of the ANALOG I/O port? The Standby line is active. STAT:COND:CMD? This command reads the status of the last command (refer to Table C-2). Example: STAT:COND:CMD?<CR> 0<CR><LF> The last command was successfully completed. STAT:COND:EVEN? This command reads system events status (refer to Table C-2). Example: STAT:COND:EVEN?<CR> 4<CR><LF> B-38 Keyswitch is in OFF position.