READ:QSW:CNTS?
This command reads the pulse counter in megacounts.
Unit: Mcnts
Examples:
READ:QSW.CNTS?<CR>
34.4Mcnts<CR><LF>
How many pulses have there been since the last pulse count reset?
The system has issued 34,400,000 pulses.
READ:SHG:TEMP?
READ:THG:TEMP?
These queries retun the measured SHG or THG temperature in counts.
Unit: cnts (counts)
Example:
READ:SHG:TEMP?<CR>
1649cnts<CR><LF>
What is the actual SHG temperature?
The actual SGH temperature is 1649 counts.
STAT:APORT:SIGNAL?
This query returns the status of the ANALOG I/O interface pins. Use this query to find the activity
level of the individual signals applied to the ANALOG I/O port (refer to Table B-5).
Individual bit position = 1: signal on pin is active
Individual bit position = 0: signal on pin is not active
Example:
STAT:APORT:SIGNAL?<CR>
64<CR><LF>
What is the status of the lines of the ANALOG I/O port?
The Standby line is active.
STAT:COND:CMD?
This command reads the status of the last command (refer to Table C-2).
Example:
STAT:COND:CMD?<CR>
0<CR><LF>
The last command was successfully completed.
STAT:COND:EVEN?
This command reads system events status (refer to Table C-2).
Example:
STAT:COND:EVEN?<CR>
4<CR><LF>
B-38