Download SC430 User Manual-AV-S
Transcript
WARNINGS AND CAUTIONS - Oxygen Monitor Oxygen Monitor NOTES WARNINGS 1. We recommend calibration of the oxygen monitor every time the system is turned on, as a safety precaution. 2. Do not attempt to open the fuel cell. The sensor contains small quantities of : a) electrolyte, classified as a harmful irritant which is potentially hazardous, and b) lead. Used or defective cells must be disposed of according to hospital, local, state, and federal regulations. 3. ALWAYS check the integrity of the sensor assembly before use. 4. Once exhausted, the sensor must be disposed of according to hospital, local, state and federal regulations. 5. The sensor measures oxygen partial pressure, and its output will rise and fall due to pressure change. An increase in pressure of 10% at the sensor inlet will produce a 10% increase in sensor output. CAUTIONS 1. Only use low temperature ethylene oxide sterilisation for the oxygen sensor. The sensor is not compatible with other sterilisation techniques - damage may result. Do not sterilise any other components. 2. Do not autoclave or expose the sensor to high temperatures. 3. If the sensor shows signs of being affected by condensation, dry the sensor with soft tissue. Do not use heat to dry the sensor. 6 1. The O2 SENSOR FAULT alarm indicates that one of the following conditions has occurred. a) Internal electrical fault b) Software/electronics fault c) Oxygen sensor fault. 2. The concentration read-out may, in certain conditions of excess pressure, show a value above 100%. To accommodate these conditions it is possible to set the high alarm value up to 105% (see section 5). 3. To maintain maximum sensor life: i) always switch off the anaesthetic machine after use, to ensure that the basal flow ceases. ii) disconnect the breathing circuit after use. 4. The accuracy of flow and volume measurements may be reduced if the oxygen monitor is not in use. 5. Fresh gas mixture compensation is disabled if the oxygen monitor is switched OFF.