Download here - Hach

Transcript
Standard programs
Enter the Variables (Factors, Wavelength and Coversion
Factors)
For detailed information on entering the Variables, refer to
Section 6.1.2, page 82.
Check the process specification in advance to determine whether
this point needs to be changed.
1. Highlight the Variables line in the data overview and press
Edit.
2. Select the variable; press Edit, and enter the data (refer to
process specification) using the alphanumeric keypad (for
F1, F2, 11, U1 etc.). Confirm each entry with OK.
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
1: Wavelength 1
U1: Conversion factor 1 for the first chemical form
U2: Conversion factor 2 for additional chemical forms, etc.
.
Timer 1, Timer 2, Timer 3 and Timer 4:
The timers are activated or deactivated with the control boxes on
the left of the display. In the next column, a selection can be
made from a list of names that describe the corresponding work
step. In the third column, the times for each active timer are
entered.
3. When finished programming, press Store and then Cancel
to return to the "User Programs" menu.
4. Press Start to begin the test with the new data.
69