RSS-131-OM User’s Manual Rev. R
Diagnostic Commands
There are a number of serial commands that are only used for internal GE
diagnostic purposes. They are not intended for customer use. These commands
are documented only for informational purposes.
Command
B
J
Y
Function
Display A/D data
Display queue data as it is written for one input
Reboot device
Table 6: Diagnostic Serial Commands
Display Raw A/D Data (#B)
This command will display raw A/D data for all channels and the electrometer
range. One line will be sent every time new data is available, which occurs once
per second. It will continue to do this until any character is received at the COM
port. The A/D values are the raw values the micro controller reads at its input port
connected to the Analog to Digital converter. A table of the A/D inputs and their
functions are described in the following table. The wind speed, electrometer
ranges and rain gauge are not actually A/D inputs, but are included for
completeness of sensor data.
Column
Sensor
0
Wind speed
1
Unused
2
Barometric pressure
3
Wind direction
4
HPIC (radiation)
5
High voltage bias
6
Battery
7
Temperature
8
Electrometer range
9
Rain gauge
COMMAND
#B yy b<CR>
yy = the HPIC ID. A one byte value that ranges from ‘0’ to ‘Z’ (ASCII) excluding
‘@’.
RESPONSE
115