Transcript
PVCHECK modules’ Voc. “Manual” setting, made by the operator, of the known temperature value of the module in the line below. Aux temperature measurement with auxiliary probe PT1000. Tol. Voc (%) percentage value of the desired limit tolerance (set by the operator according to his/her own requirements) for the measurement of Voc carried out by the instrument. Allowable values: 0% 25%. The value between brackets (4%) indicates the reading error of the instrument in Voc measurement. Tol. Isc (%) percentage value of the desired limit tolerance (set by the operator according to his/her own requirements) for the measurement of Isc carried out by the instrument. Allowable values: 0% 25%. The value between brackets (4%) indicates the reading error of the instrument in Isc measurement. Ins Test enabling/disabling of insulation measurement and setting of test voltage. Possible options: OFF, 250V, 500V, 1000V. With enabled function, the line “Ri min” appears, which allows setting the minimum limit threshold in the range 0.1 100M RPE Test enabling/disabling of continuity test and setting of the limit threshold value for measurement. Possible options OFF, 1 5 in steps of 1 5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen. CAUTION The settings saved for test voltage for the insulation measurement carried out within the IVCK function also affect the settings for the single measurement (MENU M). EN - 19